3- SAFETY RULES FOR INSTALLATION
DANGER! Tecnomed Italia s.r.l. declines any liability, expressed or implied, and cannot be
held liable for personal injury and/or direct or indirect property damage deriving from failure to
comply with the instructions in this manual and/or from incorrect installation and/or use of the
device and its accessories and/or from improper and/or lack of cleaning and/or maintenance.
ATTENTION! For safety reasons, Tecnomed Italia s.r.l., manufacturer of the Morpheus
surgical chair, recommends that the installation, maintenance and repair operations are
carried out only by authorised technical staff of Tecnomed Italia s.r.l.
DANGER! In case of faults, the components are to be replaced using only original spare parts.
DANGER! Any technician not authorised by Tecnomed Italia s.r.l., who makes changes to
the product replacing parts or components with spare parts different from those recommended
by the manufacturer, assumes a responsibility similar to that of the manufacturer.
DANGER! It is strictly forbidden to make changes to the device. Tecnomed Italia s.r.l.
assumes NO responsibility for personal injury and/or direct or indirect property damage
resulting from non-compliance with this requirement.
ATTENTION! The Morpheus surgical chair can be either fixed or mobile.
DANGER! Comply with the conditions of use provided in the “Technical Specifications”
chapter and do not exceed the recommended values.
ATTENTION! Before connecting the power cord, make sure the electrical contacts are
perfectly dry. If necessary, use compressed air to dry them.
DANGER! The electrical system available at the installation site must comply with
standards CEI 64-8/7 and subsequent amendments. The device shall be installed only by
technical staff authorised by Tecnomed Italia s.r.l.
DANGER! To avoid the risk of electric shock, this device must be connected to an
earthed electrical system.
ATTENTION! Make sure the mains voltage matches the power supply voltage specified
on the label .
DANGER! Do not install the device in environments subject to anaesthetic or flammable gases.
ATTENTION! Do not expose the device to direct sunlight or to sources of UV light.
DANGER! Install the device in a place protected against collisions or against accidental
spills of water or liquids.
DANGER! Do not install the device next to heat sources. Install it so as to ensure proper
ventilation inside it.
DANGER! Do not use the product with accessories not approved by Tecnomed Italia
s.r.l. and without CE marking or fitted with standardised interfaces. The patient and the dentist
might get injured or the device might get damaged. Install only accessories authorised by
Tecnomed Italia s.r.l. The company assumes NO responsibility for personal injury and/or direct
Содержание MORPHEUS
Страница 1: ...V 8 00 CODE DE1000X Installation use and maintenance manual MORPHEUS surgical chair ...
Страница 14: ...14 ENGLISH 5 3 1 DIMENSIONS ...
Страница 19: ...19 ENGLISH Positioning examples for standard configuration ...
Страница 20: ...20 ENGLISH ...
Страница 24: ...24 ENGLISH 7 8 9 10 11 12 ...
Страница 25: ...25 ENGLISH 13 14 15 X4 16 17 18 ...
Страница 47: ...47 ENGLISH 16 ELECTRICAL TEST Electrical safety test of earth ground and electrical insulation ...