37
ENGLISH
9 - MAINTENANCE
DANGER! Cut off the power supply to the equipment before carrying out any main-
tenance operation, by pressing the green ON/OFF light button placed on the base of the
surgical chair.
DANGER! The product requires technical safety checks and extraordinary mainte-
nance every 2 years, to identify any electrical faults (such as damaged cable insulations).
These checks are to be carried out by an authorised Tecnomed Italia s.r.l. technician.
ATTENTION! The staff of the practice is responsible for making sure that the product is
serviced in due time. Tecnomed Italia s.r.l. denies any liability, expressed or implied, and
cannot be held liable for injuries to persons and/or direct or indirect damages to property,
deriving from failure to perform the safety technical controls and maintenance operations.
The product can be used only if all the technical safety controls had positive result.
DANGER! The technical controls must also be carried out after making changes to the
machine; for example, when installing new components to avoid any problems that might
affect the electrical safety of the equipment.