background image

IST-2182.PM01.01 

Istruzione / User’s Manual / Manuel d’utilisation 

Pag. 6 / 10 

 

 

 

TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI ) 

Tel. +39 02 26922890 - Fax +39 02 2133734

 

 

Calibration Kit, Sample Gas Bottles

 (

for Calibration Check and Full Calibration

), only one of these can be used: 

20%LEL (0.88%v/v)

 

Methane in Air (20.9% Oxygen).

 

50%LEL (2.2%v/v)

 

Methane in Air (20.9% Oxygen).

 

Catalytic sensors  cannot work in oxygen deficiency

. It  is possible  to use  either the  disposable cylinders with 

adjust valve or the high pressure ones with reduction gear.  It is also necessary to use 

the calibration kit 

Tecno-

control model TC011 

(

for non-corrosive gases

). 

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the equipment. After having put the system in safety and inserted the "

Code Test

". 

The check lasts a total time of about 16 seconds, then the detector will return to normal operating conditions. It is 

advisable to perform this operation every 6-12 months according to the use. 

Note: 

The operation can not be performed if the output is greater than 

5.6mA (10% LFL)

STATUS of 

OPERATION CHECK

 

YELLOW LED GREEN LED  mA OUTPUT (S) 

Control Unit 

After entering the Check Code

 

OFF 

OFF 

0 mA 

FAULT 

Starting Operation Check

 

ON  

ON 

0 mA 

for 6 seconds

 

FAULT 

Within 5 sec. the mA output  

ON 

ON 

increases to 4 mA for 7 seconds

 

NORM 

Within 5 sec. the output in mA 

ON 

ON 

increases to 

20 mA

 for 7 seconds

 

ALARM 

End Check and return to normal operation 

OFF 

ON 

gas measurement 

NORM 

“ZERO ADJUST” (

Zero Code

: F2, F1, F1, F2)

even if there is an automatic zero adjustment system (zero tracker) 

this operation 

must be performed only if the output is different from 4mA

 (0% LFL on the control unit) after the 

installation and every 6-12 months based on the environmental conditions. This function must be done in clean air 
only (

environment without the presence of gas or other pollutants

). Immediately after having inserted the "

Zero Code

", as a con-

firmation of the operation carried out there will be 

1

 flash of the Green Led and the output will become 4,0mA. 

Note: 

The operation can not be performed if the output is greater than 

5.6mA (10% LFL)

.

 In this case, it will be 

necessary to recalibrate the unit and/or replacing the detector.

 

“CALIBRATION CHECK” (

Cal Check Code

:  F2, F1, F2, F1)

:

 

this  operation  allow  to  effect  a  real  functional  test  of 

the equipment with gas

 

after the "

Calibration

" routine, or after the installation. The “

Calibration Check

” routine should 

be done during the periodic maintenances because this is the only method to verify the instrument real functioning. 

WARNING:  the  Cal  Check  Code  allows  to  use  the  same  cylinder  Methane  +  Air  for  all  types  even  if  they 

have different gas response.

 

Without the Verification Code, using e.g. a cylinder 0.88% v/v Air mixture (corresponding to 20% LFL Methane), to 
check the SE182PG, it would be obtained that: 20% LFL Methane multiplied by 1.55 K Butane, (

see Table 2

), would become 

31%LFL = 8.96 mA. If instead, e.g. to check the SE182PG, a cylinder with a 0.28% v/v Air mixture (corresponding to 
20% LFL Butane) is used, the code must not be used and the result will be 20%LFL = 7.2 mA. 

1) 

With the keys perform the “

Cal Check Code

 ". Wait until the Yellow Led starts to flash (the green one remain fix) 

2) 

 Insert  the  TC011  over  the  sensor  holder.  Adjust  the  sample  gas  bottle  valve,  as  the  flow  meter  indicates 
around 0.3 l/mins (

see Fig.2

). 

3) 

Verify with a voltmeter connected to the Test-Point (

see Fig.1 and Table 3

), reaches the indicated value (

e.g. the 

mA output increases to the corresponding value and the control unit, to which the detector is connected, display 
the value in %LFL

). If the value result is more different, it is necessary to recalibrate the sensor (

see “

Calibration”

). 

Used Sample Gas Bottles 

Test-Point (TP1-TP2) 

mA Output (S) 

% LFL 

20%LFL (0.88%v/v) Methane in Air (20.9% Oxygen) 

69÷75 mV 

7,2mA (± 0,3) 

20%LFL (± 2) 

50%LFL (2.2%v/v) Methane in Air (20.9% Oxygen) 

114÷126 mV 

12 mA (± 0,6) 

50%LFL (± 4) 

4) 

Then, close the gas bottle, remove TC011, 

press the key F2 to restore the normal working conditions

 and 

the mA output will slowly decrease up to 4mA. 

WARNING:  For  safety,  five  minutes  after  insertion  of  the  code,  the  detector  returns  to  normal  operation. 
Remember that during the verification of the Central Unit will activate alarm devices. 

“FULL CALIBRATION” (

there are two different available Codes. Calibration 20% and Calibration 50%

)

this function allows 

completely recalibrate the sensor and must be performed when the "

CALIBRATION CHECK

" has not been exceeded. 

Important note:

 

During Calibration routine the mA output indicates 0mA

WARNING

:  "

Full  Calibration

"  should  be  performed  only  in  clean  air  (environment  without  presence  of  flammable 

gases or other pollutants) Depending on which cylinder was chosen, only the corresponding code should be used, 
in case of error the procedure will not be performed. 

1.3. 

If you have chosen to use the cylinder with 

20% LFL (0.88%v/v) Methane Gas mixture in air (

20.9 about% O

2

must be performed the 

“20% Calibration Code: F2, F2, F2, F1, F2, F1

” 

1.4. 

If you have chosen to use the cylinder with 

50% LFL (2.2%v/v) Methane Gas mixture in air (

20.9 about% O

2

must be performed the 

“50% Calibration Code: F2, F1, F1, F1, F2, F2, F1

” 

6. 

After executing the code, the Green LED will start to flash. Wait for the Yellow LED to light on. 

7. 

Insert the TC011 on the sensor holder (hold it down) and adjust the sample gas bottle valve, as the flow me-
ter indicates around 0.3 l/mins (

see Fig.2

). 

8. 

Wait around 3 minutes, then when the Green Led switches on (

and while it's 

SWITCHED ON

), press the key F2 on 

the instrument and hold it pressed until the two LEDs are switched off. (

WARNING

: if Green continuous flash-

ing over 5/6 minutes calibration has failed. In this case, turn the detector off and on, after preheating, repeat 
the  "

Calibration

"  procedure.  Verify  that  the  problem  does  not  depend  on  the  gas  or  the  unstable  flow.  If  the 

condition persists, it will be necessary to replace the detector and/or send it to the supplier for repair.

). 

Содержание SE182PB

Страница 1: ... jusqu a 50 LFL LII Limite del Sensore Sensor limits Limite échelle 100 LFL LII Sensore Sensor Capteur Catalitico Catalityc Catalytiques Preriscaldo Preheating Préchauffage 30 secondi seconds secondes Tempo di risposta T50 Response time T50 Temps de résponse T50 30 secondi seconds secondes Precisione Accuracy Precision 10 Deriva a lungo termine in aria pulita Long time drift in pure air Dérive à l...

Страница 2: ...ocedere alla Re golazione del 4mA come indicato avanti in VERIFICA FUNZIONAMENTO Funzionamento Normale deve essere acceso il solo Led Verde ON Guasti in caso di guasto del sensore l uscita S va a 0 mA FAULT Il segnale sarà poi interpretato dalla centrale e segnalato come una situazione di guasto Il Led Giallo si accende ogni 4 secondi con il Led Verde acceso per avvisare che il Sensore ha superato...

Страница 3: ... Ossigeno circa Gas Metano al 50 LFL 2 2 v v in aria 20 9 Ossigeno circa I sensori Catalitici non possono assolutamente funzionare in carenza d ossigeno È possibile usare sia bombole monouso con valvola d erogazione sia ricaricabili ad alta pressione con riduttore di pressione Inoltre è necessario il kit di calibrazione Tecnocontrol mod TC011 per gas non corrosivi TEST ELETTRICO Codice Test F2 F2 ...

Страница 4: ...icamente il rilevato re si spegne e si riavvia in funzionamento normale vedi capitolo FUNZIONAMENTO Preriscaldo Solo Led Giallo Acceso la calibrazione è fallita In questo caso dopo 8 secondi il rilevatore si riavvia automati camente e dopo il preriscaldo ripetere la procedura di Calibrazione senza reinserire il Codice Se la condizione persiste sarà necessario sostituire il rilevatore e o inviarlo ...

Страница 5: ... pole non reversible and plug in terminals For con nections and distance see also the specific instructions of the control unit to which the detector must be connect ed We recommend using a shielded 3 wire cable with the shield connected to the earth on the control unit side The maximum distance from the control unit to which the detectors can be connected depends on the power supply See Table 1 N...

Страница 6: ...iplied by 1 55 K Butane see Table 2 would become 31 LFL 8 96 mA If instead e g to check the SE182PG a cylinder with a 0 28 v v Butane Air mixture corresponding to 20 LFL Butane is used the code must not be used and the result will be 20 LFL 7 2 mA 1 With the keys perform the Cal Check Code Wait until the Yellow Led starts to flash the green one remain fix 2 Insert the TC011 over the sensor holder ...

Страница 7: ...ever ma peut détecter en même temps d autres gaz combustibles présent dans la pièce Sur le circuit situé à l intérieur du couvercle du boîtier se trouvent les touches de codage F1 et F2 pour les opéra tions de vérification et calibrage et 2 LED indiquant les conditions de fonctionnement Led Vert ON fonctionnement normal Led Jaune FAULT capteur en panne ou déconnecté ou saturé ou échu Préchauffage ...

Страница 8: ...Les autres substances causant une perte tempo raire de sensibilité sont les halogènes l hydrogène sulfuré le chlore les hydrocarbures chlorés trichloréthylène ou tétrachlorures de carbone Après une brève période en air propre le capteur reprend son fonctionnement normal VERIFICATIONS PERIODIQUES Remarque Important les opérations suivantes doivent être exécutées seulement par personnel expert et fo...

Страница 9: ...VERTISSEMENT Durant l opération d étalonnage l appareil porte la sortie S à 0mA L étalonnage doit être exé cuté en air propre exclusivement milieu sans la présence de gaz inflammables ou autres En fonction du cylindre choisi seul le code correspondant doit être utilisé En cas d erreur la procédure ne sera pas effectuée 1 1 Si la bouteille est utilisée avec un mélange de Methane à 20 LII 0 88 v v e...

Страница 10: ...s heavy gas densité par rapport à l air gaz léger gaz lourd 3 Valore in di volume del 20 LFL del gas rilevato Value in volume of 20 LFL of the gas detected Valeur en du volume correspondant à 20 de la LII du gaz détecté 4 Coefficiente di sensibilità riferito al gas Metano Gain with respect to Methane Coefficient de sensibilité par rapport au gaz methane Tabella 3 Table 3 Tableau 3 SE182P Uscita Ou...

Отзывы: