Page
9
of
52
GENERAL
MENU
Ref.
Parameter
Name
Min
Max
Default
Menu
3131
Menu
3211
#1.01
Buzzer
at
end
of
cycle
0=
Off
1=
On
On
#1.02
Alarm
buzzer
0=
Off
1=
On
On
#1.03
Key
press
buzzer
0=
Off
1=
On
On
#1.04
External
device
1
0=Off
1=Printer
2=
LAN
Off
#1.05
External
device
2
0=
Off
1=
USB
2=Bluetooth
USB
#1.06
Demineralised
water
0=
Off
1=
On
On
#1.07
Hot
water
0=
Off
1=
On
Off
#1.08
Operator
PIN
0=
Off
1=
On
Off
#1.09
Door
unlocked
at
end
of
cycle
0=
Off
1=
On
Off
#1.10
Font
Display
0=
LCD
(EUR
‐
JAP)
1=
LCD
CY
(EUR
‐
CYR)
LCD
EUR
‐
JAP
#1.11
Language
selection
1
8
2
#1.12
Restart
mode
after
alarm
0
2
2
#1.13
Key
1
program
1
40
1
#1.14
Key
2
program
1
40
2
#1.15
Key
3
program
1
40
3
#1.16
Printer
language
1
8
2
#1.17
Program
to
be
repeated
1
40
1
#1.18
Number
of
repetitions
0
99
0
#1.19
External
devices
mode
0=
Not
locking
1=
Locking
Not
locking
#1.20
Microswitch
board
2
enabling
0=
Off
1=
On
Off
#1.21
Drying
enabling
0=
Off
1=
On
On
#1.22
Spray
arm
control
enabling
0=
Off
1=
On
Off
#1.23
White
LED
light
always
on
0=
Off
1=
On
Off
#1.24
Salt
reserve
sensor
enabling
0=
Off
1=
On
On
#1.25
Number
of
regenerations
for
salt
reserve
1
99
3
TEMPERATURE
MENU
Ref.
Parameter
Name
Min
Max
Default
Menu
3131
Menu
3211
#2.01
Hysteresis
0
99
2
°C
#2.02
Maximum
pre
‐
wash
temperature
0
99
35
°C
#2.03
Maximum
tank
temperature
0
250
97
°C
#2.04
Maximum
air
temperature
0
250
150
°C
#2.05
Maximum
temperature
difference
between
the
probes
in
the
tank
0
99
3
°C
#2.06
Offset
beyond
the
phase
temperature
0
99
10
°C
#2.07
Minimum
difference
control
temperature
of
probes
in
tank
0
99
80
°C
#2.08
Minimum
drying
temperature
0
250
120
°C
#2.09
Cooling
fan
activation
temperature
0
100
50
°C
Содержание TD010TMD
Страница 1: ...ENGLISH Service manual THERMODISINFECTOR TD010TMD TD011TMD TD012TMD Service manual 0TMDT0002 REV 0...
Страница 47: ...Page 47 of 52 DIAGRAMS...
Страница 48: ...Page 48 of 52...
Страница 49: ...Page 49 of 52 TD010TMD TD012TMD...
Страница 50: ...Page 50 of 52 TD011TMD...