E
TITLE
DESCRIPTION
Device
Possible
causes
Solutions
45
Tank
T.
limit
During
the
washing
phase,
a
temperature
in
the
tank
higher
than
the
one
set
has
been
detected.
1.
Temperature
increase
above
setpoint.
1.
The
8KM1
contactor
of
the
tank
heating
element
is
stuck
PT1000
11ST1
or
11ST3
temperature
probe
not
in
housing
or
faulty
46
Air
probe
T.
limit
During
the
washing
phase,
a
temperature
higher
than
the
maximum
one
set
has
been
detected.
2.
Temperature
increase
above
setpoint.
2.
The
contactor
of
the
8KM2
air
heating
element
is
stuck
PT1000
11ST2
temperature
probe
not
in
housing
or
faulty