5.2
Chemicals
calibration:
The
following
procedure
must
be
carried
out
after
completing
the
operations
listed
previously
in
paragraph
“3.1
Preparation
of
chemicals
for
calibration”:
1.
Press
PRG
to
return
to
the
menu;
2.
Select
"chemicals
calibration"
and
press
START;
3.
Select
the
chemical
to
be
calibrated,
browsing
with
keys
1
and
2
and
making
sure
that
the
silicone
pipe
is
inside
the
beaker;
4.
Press
START
to
start
pouring
the
liquid
in
the
beaker;
5.
Press
RESET
when
the
liquid
reaches
100
ml
inside
the
beaker
to
stop
the
liquid
from
being
poured;
6.
Press
PRG
to
go
back
to
the
previous
menu.
This
way,
the
calibration
just
carried
out
will
be
saved
in
the
memory;
7.
Now
the
calibration
of
the
product
concerned
is
complete.
6.
MESSAGE
MENU
SETTINGS
There
are
two
USB
sockets
for
machine
programming
on
the
side
of
the
machine.
The
USB
socket
on
the
left
is
used
to
connect
a
PC
and
to
use
the
“Monitor
WD2”
program;
the
USB
slot
on
the
right
is
used
to
transfer
precompiled
programs
from/to
a
USB
flash
drive.