57
DXBA408
Rev_14
Warmig up time and vacuum fractionation can vary, because of autoclave’s conditions, between 25 and 30
minutes. You have to an this time to sterilization’s times showed in table
9.4 – TEST CYCLES
-Helix test: this test serves to verify steam penetration capacity in a hollow body. Use the Helix Indicator
System.
- Bowie and dick test: this serves to verify steam penetration capacity in a porous load. Use Bowie and Dick
test pack (top table).
- Vacuum test: this serves to verify the preservation capacity of the autoclave vacuum; test report is prepared
by the autoclave printer.
It is preferable to use tests distributed by TECNO GAZ spa.
TEST CYCLES TABLE
CYCLES
STERILIZATION
TIME
( Minutes)
DRYING TIME
(Minutes)
MAX. LOADING FOR
TRAY
(Kg)
MINIMUM WORK PRESSURE
( bar, differential pressure )
134°C Helix/
Bowie&Dick
3.5 4
2.02
134°C Vacuum
-0.9
CYCLE DIAGRAM
Содержание Europa B xp
Страница 31: ...31 DXBA408 Rev_14 SCHEMA DI COLLEGAMENTO DEL DEMINERALIZZATORE VALIDO FINO AL 11 05...
Страница 32: ...32 DXBA408 Rev_14 SCHEMA DI COLLEGAMENTO DEL DEMINERALIZZATORE VALIDO DAL 12 05...
Страница 67: ...67 DXBA408 Rev_14 CHEME OF CONNECTION OF THE VALID DEMINERALIZER UNTIL THE 11 05...
Страница 68: ...68 DXBA408 Rev_14 SCHEME OF CONNECTION OF THE VALID DEMINERALIZER FROM THE 12 05...
Страница 104: ...104 DXBA408 Rev_14 SCHEMA DE BRANCHEMENT DU DEIONISATEUR VALIDE JUSQU AU 11 05...
Страница 105: ...105 DXBA408 Rev_14 SCHEMA DE BRANCHEMENT DU DEIONISATEUR VALIDE DU 12 05...
Страница 141: ...141 DXBA408 Rev_14 ESQUEMA DE ENLACES DEL DESMINERALIZADOR V LIDO HASTA EL 11 05...
Страница 142: ...142 DXBA408 Rev_14 ESQUEMA DE ENLACES DEL DESMINERALIZADOR V LIDO DEL 12 05...
Страница 178: ...178 DXBA408 Rev_14 ANSCHLUSSCHEMA DES G LTIGEN ENTZSALZERS BIS DAS 11 05...
Страница 179: ...179 DXBA408 Rev_14 ANSCHLUSSCHEMA DES G LTIGEN ENTZSALZERS VOM 12 05...