Tecno Control SE237K Скачать руководство пользователя страница 6

IST-2237.KM01.01

 

Istruzione / User’s Manual / Manuel d’utilisation 

Pag.6/10

 

 

 

 

TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI ) 

Tel. +39 02 26922890 - Fax +39 02 2133734

 

 

 

TEST and CALIBRATION 

PAY  ATTENTION

This  procedure  has  to  be  made  with  extreme  attention  and  by  authorized  and  trained 

people;  because  starting  this  procedure  it  will  start  both  Outputs  (relays)  causing  the  activation  of  con-
nected alarm devices. 

Operation Check, Zero Adjust, Calibration Check and Calibration

:

 are different code protected functions. To ac-

cess these functions is necessary to insert the relevant “

Code

" through the keys F1s and F2. To have the key pres-

sure recognized, hold pressing it for around a second (

until the Green Led doesn't switch off for a moment

). Then the next key 

can be pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automatically erased. 

Calibration Kit, Sample Gas Bottles

 (

for Calibration Check and Calibration

) please, only using a mixture: 

20%LEL (0.88%v/v)

 

Methane in Air (20.9% Oxygen).

 

Catalytic  sensors  cannot work in oxygen deficiency

. It  is possible  to use  either the  disposable cylinders with 

adjust valve or the high pressure ones with reduction gear. It is also necessary to use 

the calibration kit 

Tecno-

control model TC011 

(

for non-corrosive gases

) or 

TC014

 

(

Stainless steel

). 

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the equipment. After having put the system in safety and inserted the "

Code Test

", all Led are switched off and all 

the  relays  are  disarmed.  Then  they  will  switch  on  in  sequence,  the  Led,  from  the  yellow  up  to  the  3°  red.  To  the 
lighting  of  the  various  Led  it  corresponds  the  activation  of  the  relevant  relays  (

relay  "FAULT"  with  the  yellow  Led,  relay 

"ALARM1" with the 1° red Led, relay "ALARM2" with the 2° red Led, relay "ALARM3" with the 3° red Led

). At the end all the Led will 

remain lighted for around 5 seconds, then the central returns at the conditions of normal operation. It is advisable to 
perform this operation every 6-12 months according to the use. 

Note: this function is not working if the 1° and/or the 2° and/or the 3° red Led have already turned on 

 

“ZERO ADJUST” (

Zero Code

: F2, F1, F1, F2)

this function is to adjust the Zero sensor and can be done in clean 

air only (environment without the presence of gas or other pollutants). Immediately after having inserted the "

Zero 

Code

", as a confirmation of the operation carried out there will be 1 flash of the 1° red Led and the output will be-

come 4,0 mA.  

We suggest performing this operation 

only if the 1st Red LED is on

 (or output is different from 4mA) 

after the in-

stallation or after the change of the cartridge and every 6-12 months based on the environmental conditions. 

Note

This function is not working if the 2° e/o the 3° red Led have already been switched on .In case the 1° 

red Led is already switched on only instead of flashing it will switch off for confirmation.

 

“CALIBRATION” (

Calibration Code

: F2, F2, F2, F1, F2, F1)

this function allows completely recalibrate the sensor.

 

Warning

: to guarantee that no errors of elaboration happen, the rare possibility exists that during the Calibration 

the 

yellow LED switch off every 8 seconds, 

in this case interrupt the procedure, switch off and witch on the instru-

ment and repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the 
reparation. 

Important note:

 

During Calibration routine the mA output indicates 0mA

.

 

The Calibration can be done in clean air only (

environment without the presence of flammable or other polluting gas

). With the 

keys perform the “

Calibration

 Code

 ". Wait until the Yellow and Green Led switch on fix and the 1° red Led starts to 

flash. Insert the TC011 into the sensor holder to regulate the influx of the gas (

0.88%v/v methane in air

), the way that 

the flow meter indicates around 0.3 l/mins (

see Fig.3

). Wait around 3 minutes, then when the 3° red Led switch on 

(

and  while  it's  SWITCHED  ON

),  press  the  key  F2  on  the  instrument  and  hold  it  pressed  until  1°  and  3°  Red  Led  are 

switched off for at least 2 seconds (

if the 1° Red Led continue to flash, wait that the 3° Red Led switch on again and repeat the op-

eration

). 

Then, close the gas cylinder and remove TC011

. At this point we can have two possibilities: 

Yellow  and  Green  Led  illuminates:

  the  calibration  routine  has  correctly  been  performed. 

Wait  8  seconds,  until 

the instrument automatically restores the normal working conditions

(

see “Operational Description > Preheating”

Yellow LED illuminates:

 the routine has failed. In this case, wait 8 seconds, until the instrument automatically re-

peat 

Preheating

, then repeat the “

Calibration”

 routine without inserting again the code. If condition still persists after the 

replacement of the cartridge, it will be necessary to send the detector back to the manufacturer for reparation. 

“CALIBRATION CHECK” (

Cal Check Code

: F2, F1, F2, F1)

:

 

this operation allow to effect a real functional test of 

the equipment with gas

 

after the "

Calibration

" routine, or after the installation. The “

Calibration Check

” routine should 

be done during the periodic maintenances because this is the only method to verify the instrument real functioning. 
With the keys perform the "

Cal Check Code

 ". Wait until the Yellow Led starts to flash (the green one remain fix) In-

sert the TC011 over the sensor holder, adjust the sample gas bottle valve (

0,88%v/v methane in air

), as the flow meter 

indicates around 0.3 l/mins (

see Fig.3

), wait for 3 minutes and check that the instrument activates Alarms as shown 

in  chapter  “

Operational  Description

”. 

If  the  result  is  different,  is  necessary  to  recalibrate  the  sensor  (

see  “Calibration”

).

 

Then, close the gas bottle, remove TC011, 

press the key F2 on the instrument to restore the normal working 

conditions

NOTE

:  further  to  the  above,  If  the  mA  output  is  also  used,  verify  with  a  voltmeter  connected  to  the  Test-Point 

TESTmA

”, the value reach a value between 

184 and 216mV

, [corresponding to 20mA (±1.6) output and the central 

unit should be display about 20% LEL (±2)]. Then, the mA output will slowly decrease up to 4mA. 

 

 

 

Содержание SE237K

Страница 1: ...lais 24V 1A SPST 4 20 mA Lineare Linear linéaire Campo di misura Standard Range Champ de mesure 0 20 LIE LEL Limite Scala Limits Limite échelle 25 LIE LEL 24mA Vita media in aria pulita Average Life in fresh air Vie moyenne en air pur 5 anni years ans Tempo di risposta T90 Response Time T90 Temps de réponse T90 60 secondi seconds secondes Ripetibilità Repeatability Répétitivité 5 del segnale signa...

Страница 2: ...spegne e si accende il Led verde ON che indica il normale funzionamento Dopo questo tempo il sensore è in grado di rilevare il gas ma raggiunge le condizioni di stabilità ottimali dopo circa 4 ore di funzionamento continuo Funzionamento Normale deve essere acceso il solo Led Verde ON Il 1 Led Rosso ALARM 1 si accende se la concentrazione di Gas supera la prima soglia se il gas non si riduce entro ...

Страница 3: ...are gas puri o l accendino direttamente sul sensore che potrebbe essere irrimediabilmente danneggiato ATTENZIONE Considerare che in ambienti particolarmente inquinati o con vapori di sostanze infiammabili in par ticolare i solventi la vita utile del sensore può ridursi notevolmente Alcune sostanze causano una riduzione per manente di sensibilità evitare che il sensore venga a contatto con vapori d...

Страница 4: ...aggiunta a quanto sopra se si usa anche l uscita in mA controllare anche con i puntali del voltmetro sui Test Point TESTmA vedi Fig 2 si raggiunga un valore tra 184 e 216 mV ovvero che l uscita in mA aumenti fino a circa 20 mA 1 6 e la centrale cui è collegato il Trasmettitore indichi circa 20 LIE 2 Poi l uscita tornerà progressivamente a 4 mA EN DESCRIPTION The SE237K series is a flammable gas de...

Страница 5: ...unit must be positioned vertically with the sensor downwards Models SE237KG should be fixed at 20 30 cm from the floor LPG and Petrol vapours are heavier than air Models SE237KM should be fixed at 20 30 cm from the ceiling Methane and Hydrogen are lighter than air Models SE237KI should be fixed at 20 30 cm from the ceiling Methane and Hydrogen are lighter than air Models SE237KB should be fixed at...

Страница 6: ...ALIBRATION Calibration Code F2 F2 F2 F1 F2 F1 this function allows completely recalibrate the sensor Warning to guarantee that no errors of elaboration happen the rare possibility exists that during the Calibration the yellow LED switch off every 8 seconds in this case interrupt the procedure switch off and witch on the instru ment and repeat the Calibration If condition persists it will be necess...

Страница 7: ...d autres gaz inflammables présents dans la même pièce Préchauffage à partir de la mise sous tension le détecteur commence la phase de préchauffage du capteur si gnalée par le clignotement de la LED jaune FAULT Après environ 60 secondes la LED jaune s éteint et la LED verte ON s allume indiquant le fonctionnement normal Après ce temps le capteur est apte à détecter le gaz mais il n atteint les cond...

Страница 8: ...alibrage INSTRUCTIONS La vie utile du capteur en air propre est 5 ans en moyenne Au terme de cette période indiqué par l instrument par un clignotement de la LED Jaune toutes les 4 secondes il est nécessaire de remplacer la Cartouche Capteur Vérifications Périodiques nous conseillons d exécuter tous les ans la vérification de fonctionnement du détec teur Test Électrique Réglage du Zéro et Vérifica...

Страница 9: ...Après 8 secondes le détecteur revien dra automatiquement dans les conditions de fonctionnement normal chapitre Fonctionnement Préchauffage LED Jaune allumée l étalonnage a échoué Dans ce cas après 8 secondes le détecteur reviendra automati quement dans les conditions d étalonnage répéter la procédure sans réinsérer la séquence Si la condition per siste même après la substitution de la Cartouche Ca...

Страница 10: ...rapport à l air gas leggero light gas gaz léger gas pesante heavy gas gaz lourd Fig 1 Dimensioni Size Dimensions 1 2 3 4 ON OFF Centrale rivelazione gas Gas Central Unit Centrale de détection RL Dip Switch mV Test Uscita Output Test Sortie Test 40 200mV Uscita in 4 20 mA 4 20 mA Output Sortie en 4 20 mA ALARM 1 ALARM 2 ALARM 3 FAULT SET Fig 2 Schema di collegamento Wiring diagram Schéma TESTmV TP2...