CSB CryoEM Facility TF20 Operating Instructions SEC
05/02/16
7.
Carefully place your grid in the opening on the tip. Make sure it resides in the circular recess of
the tip.
8.
With the tool, carefully bring down the clamp until it securely fastens the grid and clamp. Do
this with care so as to avoid jamming the ring onto the grid.
9.
Return the tool to its resting place. Replace the plastic covering on the tip. Turn the holder
upside down and gently tap the back (black part) a few times to verify that the clamp was
positioned properly and that the grid is mounted correctly and stays in place.
10.
Take the holder out of the plastic cover and align its pin to the 3 o’clock position.