PROVASET 2P code 2P2B00 - 2P6B00 - Instruction manual
rev. 14/02/2013
page 29
xxxxx
: Full scale range (fixed at 2 bar for
2P2B00,
fixed at 6
bar for
2P6B00,
no programming possibility)
yyyyy
: Sample pressure for gain calibration
Press the key:
[+], [-], [P
↵
]
to program the sample pressure parameter
[
â
]
to proceed to the calibration of the gain at sample
pressure
[
↑
]
or
[P
↵
]
to return to the previous steps without
the calibration.
xxxxx
: Analogue/digital conversion points
yyyyy
: Programmed sample pressure
By proceeding with care on the outer pressure source, set
the pressure on the "OUT" test connector to the
programmed sample pressure value.
Check the current value of the sample pressure by using
the outer digital pressure gauge.
Press the key:
[+]
to calibrate to the programmed sample pressure; it
stores the A/D conversion value like calibration constant.
It returns to the procedure beginning.
[-]
or
[P
↵
]
to return to the previous steps without the
calibration.
CALIBRATION
FULL SCALE
mmH2O XXXXX
CALIBRATE GAIN
REFERENCE:
mmH2O YYYYY
CALIBRATION
A/D XXXXX
mmH2O YYYYY
CALIBRATE !
+YES -NO