New Wellness Expert: Service & Maintenance manual - rev. 1.1
7. The operation will be repeated five times, with the target in each of the four corners and finally
in the middle of the screen.
8. At the end of the procedure, the following message will appear for a few seconds:
Calibration accepted
Error in pixel = X
If X is less than 10, as specified in the previous setting, or:
Calibration rejected
Error in pixel = X
If X is greater than 10, in which case the calibration carried out will be ignored.
8.1.2. T
OUCH SCREEN SECOND MODEL
“HAMPSHIRE”
To access the touch screen calibration function, proceed as follows:
1. Click
START
,
PROGRAMS
, then start the program “
Hampshire TSHARC Control Panel”
.
Figure 8.1-3
2. In the window which appears, choose the folder “
Calibration
” and run the calibration program,
pressing the point indicated by the arrows.
Page 8.3
Содержание Wellness Expert
Страница 1: ...SERVICE MAINTENANCE MANUAL REV 1 1 ...
Страница 2: ......
Страница 4: ......
Страница 10: ...New Wellness Expert Service Maintenance manual rev 1 1 Page 1 4 Page intentionally left blank ...
Страница 16: ...New Wellness Expert Service Maintenance manual rev 1 1 Page 2 6 Page intentionally left blank ...
Страница 20: ...New Wellness Expert Service Maintenance manual rev 1 1 Page 3 4 Page intentionally left blank ...
Страница 22: ...New Wellness Expert Service Maintenance manual rev 1 1 Page 4 2 Page intentionally left blank ...
Страница 56: ...New Wellness Expert Service Maintenance manual rev 1 1 Page 8 8 Page intentionally left blank ...
Страница 60: ...New Wellness Expert Service Maintenance manual rev 1 1 Page 9 4 Page intentionally left blank ...
Страница 63: ......