Siting Considerations
All the units are air cooled, air being pulled in through the front and
expelled from the rear. The ventilation grilles should never be
obstructed, therefore, as this will detract from the performance. Also
the fridge units should never be situated in an enclosed space or
near an extractor fan, this too will detract from the performance.
MINIMUM MINIMUM
150 mm (6") 150 mm (6")
Setting up the unit
1. Check the voltage marked on the unit’s rating plate against that
of the mains supply. Connect the mains cable to a suitable plug
as in the following table.
Note that the unit must be earthed to
ensure proper electrical safety.
Connections
230V
110/120V
Live
brown
black
Neutral
blue
white
Earth
yellow and green
green
If a fused plug is used it should be fitted with the following value
fuse:
Supply
230V
Fuse Rating
13 Amp
Note that units marked 230V on the rating plate work equally well
at 220V; units marked 120V on the rating plate work equally well
at 110V