32
4
Press the ENT button to confirm the selection.
5
Press the ESC button to exit setting mode.
ATTENTION
o
The ZERO button function depends on the Cal. Value Lock
setting.
When Cal. Value Lock is ON, press the ZERO button to use
the Digital Zero function. When Cal. Value Lock is OFF, press
it to use the Zero Balancing function.
o
In order to prevent accidental changes to the calibration
value, set Cal. Value Lock to ON after calibration.
NOTE
See “4-10 Setting value list” on page 27 for information about
settings that cannot be changed when Cal. Value Lock is ON.
5-2. Equivalent input calibration
This method determines the calibration value by recording
rated output and rated capacity values from a test results table.
Use this to calibrate easily when an actual load cannot be
applied.
An overview of the equivalent input calibration procedures is
shown below.
Sensor value memory
selection
c
Rated Output/Rated
Capacity
c
Zero Balancing
c
Select EU
c
Cal. Value Lock setting
o
See “10-2 D/A Converter” on page 55 for information
about D/A settings.
5-2-1. Sensor value memory selection
This unit can save six types of calibration values for each con-
nected sensor.
Select the memory number for saving the calibration value.
5-2-2. Excitation Voltage
The bridge voltage of this unit is fixed at DC 2.5 V.
ATTENTION
If the bridge voltage of this unit exceeds the sensor's permit-
ted excitation voltage, the sensor could be damaged.
5. Calibration