background image

NOTE

i

408-1610

8 of 11

Rev M

Refer to the table in Figure 11 for alternate hard 

gage dimensions  for tools 69692-1 and 69693-1.

A. Gaging the Crimping Chamber Using A Plug Gage

This inspection requires the use of plug gages 
conforming to the dimensions listed in Figure 9 and 
Figure 10. Tyco Electronics does not manufacture or 
market these gages. Refer to 408-7424 for information 
concerning plug gages

To gage the crimping area(s), refer to Figure 12 and 
proceed as follows:

1. Clean oil or dirt from the crimping chamber and 
plug gage.

2. Close handles of tool until wire barrel dies are 
bottomed. Do not apply additional pressure to tool 
handles.

3. With wire barrel dies bottomed, inspect the wire 
barrel crimping chamber using the proper plug 
gage. Lift the spring-loaded locator up and hold 
gage in straight alignment with the crimping 

chamber. Carefully try to insert, without forcing, the 
GO element. See Figure 12, Detail A. The GO 
element must pass completely through the crimping 
area.

4. Try to insert the NO-GO element. The NO-GO 
element may enter partially, but must not pass 
completely through the crimping area.

5. Set insulation adjustment indicator in Position 1. 
Measure both insulation crimping chambers with the 
proper GO plug gages in the same manner as Steps 
2 and 3. See Figure 12, Detail B.

6. Set insulation adjustment indicator in Position 4. 
Measure both insulation crimping chambers with the 
proper NO-GO plug gages in the same manner as 
Steps 2 and 4. See Figure 12, Detail B.

If the crimping areas conform to the gage inspection, 
the tool is considered dimensionally correct. If the 
crimping areas do not conform to the inspection, the 
tool must be repaired. Refer to Section 8, 
REPLACEMENT AND REPAIR.

Figure 9

Содержание 59275

Страница 1: ...G Radiation Resistant Terminals and Splices on stranded copper wire sizes 26 through 14 PLASTI GRIP Terminals on solid or stranded copper wire sizes 22 through 14 PLASTI GRIP Butt Splices on solid or...

Страница 2: ...408 1610 2 of 11 Rev M Figure 2 Cont d...

Страница 3: ...st fully closed Crimp the color coded terminals and splices in the matching color coded portion of the tool Refer to Section 3 3 Place terminal in crimping dies so that terminal tongue slides under lo...

Страница 4: ...sts against the recessed surface of the locator as shown in Figure 5 Spare Wire Caps Insulation Adjustment Indicator at Position 4 Wire Barrel Locator Raise Locator So the End of the Cap Rests Against...

Страница 5: ...p of tool and turn indicator to Position 4 2 Place terminal or splice in tool dies 3 Insert UNSTRIPPED wire into ONLY the insulation barrel see Figure 3 or Figure 4 of terminal or splice 4 Perform a c...

Страница 6: ...ustment indicator in Position 4 for wire having a large insulation diameter 2 Set insulation adjustment indicator in Position 3 for wire having a medium insulation diameter 3 Set insulation adjustment...

Страница 7: ...degreasing compound with a lint free cloth When degreasing compounds are not available tool may be wiped clean with a lint free cloth Relubricate tool as instructed in Paragraph 7 3 before placing it...

Страница 8: ...ring loaded locator up and hold gage in straight alignment with the crimping chamber Carefully try to insert without forcing the GO element See Figure 12 Detail A The GO element must pass completely t...

Страница 9: ...1 position 3 Insert proper solder slug into one crimp chamber see Figure 9 and Figure 10 and cycle tool 4 Measure the compressed solder slug for the wire barrel crimping chamber and check for complia...

Страница 10: ...en the bottoming surfaces of the crimping dies To inspect the CERTI CRIMP hand crimping tool ratchet control 1 Perform a crimp using the largest wire size for your tool 2 While holding the wire in pla...

Страница 11: ...r than those listed in Figure 13 should be replaced by Tyco Electronics to ensure quality and reliability of the tool Order replacement parts through your TE Representative or call 1 800 526 5142 or s...

Отзывы: