1.5.1.1 USB key
A standard USB flash memory key is used to store the results files. The unit has been tested
with keys up to 8Gb in size.
Insert the USB key into the upper USB socket on the unit before starting to test. Most USB
memory devices have an LED that indicates when the device is accessed. This will flash when
a result is written.
CAUTION Do not unplug the USB key whilst data is written
– data loss will result.
1.5.1.2 Setting the clock
The output of the unit must be switched off to set the clock.
•
Set the STORE RESULT switch to SET CLOCK.
Figure 1.18 Clock setting screen
•
Use the
and
keys to select which digit to change.
•
Use the AUX AC ADJ control to change the selected digit.
•
Use the red ON button to save the changes to the clock.
•
Use the green OFF button to cancel setting the clock.
•
Set the STORE RESULT switch to OFF.
1.5.1.3 Entering a comment
Each result is tagged with a comment when it is stored. The comment is entered by setting the
store result control to “comment” and typing the comment using the optional USB keyboard.
The same comment is tagged to each result until the comment is updated, and the comment is
erased when the unit is switched off.
Any USB keyboard can be used with the 200A-3PH mk 2.
The comment can be typed on the USB keyboard using the letters a-z, numbers 0-9 and
punctuation marks. A maximum of 39 characters may be entered for the comment.
The following keys are also used:
Shift + a-z
Enter capital letter.
Move cursor left/right.
Ins
Switch between overwrite and insert mode. By default the unit will overwrite
the character to the right of the cursor when a new character is typed.
Switching to insert mode will shift characters of the right of the cursor right
when a new character is typed.
hh:mm:ss dd-mm-yy
10:13:24 15-12-13
Содержание 200A-3PH mk2
Страница 4: ...4...
Страница 8: ......
Страница 45: ...45 Figure 5 3 Connections between units 3 WIRE SUPPLY OPTIONAL T R LINK LEAD...