122
Adjustment
This function enables the user to make adjustments to measured values to correct
inaccuracies that may occur when compared to a standard or reference temperature
measurement. By setting an adjustment equation beforehand, the adjustment function will
automatically calculate to record and display the post-adjusted measurement. It is possible
to select from two methods of adjustment: 1-point adjustment and 2-point adjustment.
*The settings can be made only via optical communication
NOTE:
When you are carrying out adjustments and formulating equations please take care with the settings and
take personal responsibility for the process.
The adjustment function can be used only with the RTR-51A, 52A, 52Pt, 53A and RVR-
52A. Note that the RTR-51, 52, 53 and RVR-52 do not contain this function.
ɡ
1-point / 2-point Adjustment
Current Reading: Enter the actual measurement before adjustment
After: Enter the desired measurement after adjustment
- Entries are valid to one decimal point.
- The input value for [Post-adjusted Reading] must be less than
Ⱦ
5 in comparison to
the input value for [Current Reading].
ɡ
Adjustment Settings
1.
Open [RTR-5W for Windows].
2.
In the Main Window, click on the Data Logger icon. To open the [Adjustment]
Window, click on [Settings] in the Menu Bar and then select [Adjustment Settings]
from the pull down menu.
[Initialize] Button
Check one
[Send Settings]
Button
NOTE (RVR-52A):
* When measuring moisture with the RVR-52A, please make adjustment settings in [Settings] – [Moisture
Probe Settings] – [Moisture Adjustment].
Under [Moisture Adjustment], check
"
Make Adjustment
"
and enter the desired values. Click the [Use
Adjustment in Equation] Button and the calculation equation will be displayed.