3 Routine Daily Inspection
The following should be checked routinely before using the tester:
Table 1: Checklist
Item Description
Appearance
Electrical connections
(Before turning the
main power on…)
Operation of each
component
·
The tester has no apparent damage.
·
The power cable is not damaged.
·
The power plug is fully inserted into the outlet.
·
The grounding cable is properly connected (only
applicable if using the ground adapter)
·
The circulating solution is filled to the specified level.
< After turning the main power on…>
·
The display appears to be functioning correctly
·
The circulating solution is circulating normally.
·
Each key functions normally when pressed.
·
The condenser bath and receiver compartment are at
the set temperatures.
·
Fire containment shutter closes correctly by pressing
Manual Fire Containment Shutter Activator. (See 10.3
on page 33)
!
Warning: Do not attempt to repair the tester yourself. In the event
that the tester exhibits any abnormal behavior
whatsoever, contact Tanaka Scientific Limited or an
authorized distributor.