- 24 -
7 Reprinting and Resending Data after Residue
After the residue is set again, corrected data can be reprinted and resent as follows:
7.1
Setting the Residue Again
Set the residue on the RUN screen. If the STANDBY screen is displayed, press the
DISPCHANGE key on the operator panel to change the display to the RUN screen (see Fig.
8). The residue can be set again by changing the value at RE CORR. To change the value,
move the cursor with the
←
or
→
key and increment or decrement the value with the
↑
or
↓
key.
7.2 Reprinting
Corrected data can be reprinted or resent only when the RUN screen is displayed.
Press the PRINT key to start printing.
Press the FUNCTION key to display the Distillation Curve screen shown in Fig.9 and the
PRINT key to start plotting a distillation curve.
7.3 Resending
Press the MANUAL CUT key to start sending the data to the RS-232C port in the same
format as at the end of a test (Note: This only applies when ADManager is OFF at Ext Out)
< TEST RESULT>
IBP
MODE(01) GASOLIN(300)
5 50.0c
70s
10
60.0c
4.5%
DIST VOL 97.5 (%)
20 75.0c 4.5%
DIST TEMP 25.0 (
o
C) 30
85.0c
4.5%
DIST TIME 300 sec
40 95.0c 4.5%
50
110.0c
4.5%
HEATER OUT 50 (V )
60 120.0c 4.5%
BATH TEMP 1.0 (
o
C) 70
130.0c
4.5%
REC TEMP 15.0 (
o
C) 80
140.0c
4.5%
HEAT TEMP 25 (
o
C) 90
150.0c
4.5%
95
170.0c
RE CORR 0.1(%)
97 185.0c
PREV.RESULT OFF
E.P 205.0c 250s
Fig.8 RUN Screen (After Test)
Содержание AD-6
Страница 2: ...1 Blank Page...
Страница 9: ...8 Blank Page...