Pro Pump Controller (ProCon-100)
Rev3.6
2023/07/17
45/70
Copyright
©
2023 by TAEHA CORP. All Rights Reserved
Table 9. Calibration Setting
Display
Function and Description
✽
Measurement
: Using this button, you can move to
‘Measurement Menu’
.
✽
Result Input
: Using this button, you can move to
‘Measurement Value Input Menu’
.
✽
EXIT
: Select this when moving to the previous
screen.
✽
Volume
: Enter the dispensing volume which you
want to measure.
✽
Speed
: Sets the dispensing speed.
✽
MESU
: Select this when dispensing is completed
after setting the dispensing condition.
✽
EXIT
: Select this when moving to the previous
screen.
✽
To calculate the measurement value, repeat several
times and input the average value.(About 10 times
recommended)
✽
After making measurements in the Measurement
Menu, input the results and save them.
✽
After this, the measured volume is automatically
compensated based on this value..