9100A/9200A/9400 

Maintenance and Performance Checks 3-5 

 

Preparation: 

1.  Configure the Oscilloscope as follows: 

Time Base: 

200 

μ

s/div 

Amplitude: 50 

V/div 

2.  Connect the 9400 Channel 1 output to the load resistance 
3.  Connect the x100 high voltage probe across the load resistance
4.  Configure the Waveform Generator as follows:  

Amplitude: 

8  V  

Output: On 
Frequency: 

As required for the tests 

5.  Connect the waveform generator to Channel 1 input. Terminate 

the waveform generator using the 50

Ω

 feedthrough terminator at 

the 9400 input  

Test Procedure 

1.  Using the variable vertical adjustment on the oscilloscope, 

adjust the vertical trace to show exactly 6 vertical divisions 

2.  Perform bandwidth, large signals tests on channel 1 output 

using Table 3-3 

3.  Use the same procedure to check channels 2, 3 and 4  

 

Table 

 

3-3, Output Bandwidth, Large Signals Tests 

Arb Frequency 

 

Oscilloscope Reading 

 

 

Setting Error 

Limits 

CH1 

CH2 

CH3 

CH4 

Pass 

Fail 

1 kHz 

6 Divisions 

 

 

 

 

 

 

100 kHz 

6 ±0.5 Divisions 

 

 

 

 

 

 

400 kHz 

6 ±1 Divisions 

 

 

 

 

 

 

550 kHz 

6 ±1.8 Divisions 

 

 

 

 

 

 

 

Bandwidth, Small 
Signals 

 

Equipment

Oscilloscope, Arbitrary Waveform Generator, Load 

Resistance, 50

Ω

 feedthrough terminator

 

Preparation: 

1.  Configure the Oscilloscope as follows: 

Time Base: 

200 

μ

s/div 

Amplitude: 5 

V/div 

2.  Connect the 9400 Channel 1 output to the Oscilloscope input. 

Attach the load resistance at the input terminals of the 
oscilloscope 

3.  Configure the Waveform Generator as follows:  

Amplitude: 

200 mV  

Output: On 
Frequency: 

As required for the tests 

4.  Connect the waveform generator to Channel 1 input. Terminate 

the waveform generator using the 50

Ω

 feedthrough terminator at 

the 9400 input  

Test Procedure 

1.  Using the variable vertical adjustment on the oscilloscope, 

adjust the vertical trace to show exactly 6 vertical divisions 

Содержание 9100A

Страница 1: ...r Electronics Ltd Tabor Electronics Ltd P O Box 404 Tel Hanan Israel 20302 Tel 972 4 821 3393 FAX 972 4 821 3388 PUBLICATION DATE January 26 2009 Copyright 2006 by Tabor Electronics Ltd Printed in Isr...

Страница 2: ...epresentatives and to users of its product and to any other person or persons for special or consequential damages of any kind and from any cause whatsoever arising out of or in any way connected with...

Страница 3: ...sion cord or a three prong two prong adapter This will defeat the pro tective feature of the third conductor in the power cord Maintenance and calibration procedures sometimes call for operation of th...

Страница 4: ...e was demonstrated to the following specifications as listed in the official Journal of the European Communities Safety IEC EN 61010 1 2nd Edition 2001 C1 C2 EMC EN 50081 1 Emissions EN 55022 Radiated...

Страница 5: ...2 1 Introduction 2 1 Conventions Used in this Manual 2 4 9100A 9200A 9400 Feature Highlights 2 5 Options 2 5 Manual Changes 2 5 Specifications 2 5 Front Panel Controls and Indicators 2 6 Mains Power...

Страница 6: ...Tests 3 3 Amplifier Bandwidth 3 4 Bandwidth Large Signals 3 4 Bandwidth Small Signals 3 4 Amplifier Pulse Response 3 5 Rise Fall Time Tests 3 5 Overshoot Tests 3 6 Amplifier Distortion 3 6 Distortion...

Страница 7: ...3 3 Output Bandwidth Large Signals Tests 3 5 Table 3 4 Output Bandwidth Small Signals Tests 3 6 Table 3 5 Rise Fall Time Tests 3 6 Table 3 6 Overshoot Tests 3 7 Table 3 7 Distortion Tests 3 8 Table 3...

Страница 8: ...iv This page intentionally left blank...

Страница 9: ...uires normal precautions and procedures applicable to handling of sensitive electronic equipment The contents of all shipping containers should be checked for included accessories and certified agains...

Страница 10: ...The 9400 is intended for indoor use and should be operated in a clean dry environment with an ambient temperature within the range of 0 C to 40 C WARNING The 9400 must not be operated in explosive dus...

Страница 11: ...r system Crest factor ratio of peak voltage to rms should be typically within the range of 1 3 to 1 6 at 10 of the nominal rms mains voltage WARNING DO not connect the line cord to the 9400 before you...

Страница 12: ...ocal Tabor representative or the Tabor Customer Service Department 1 Repack the instrument using the wrappings packing material and accessories originally shipped with the unit If the original contain...

Страница 13: ...bout 1 5 inches of ventilation space on all sides of the instrument Rack Mounting The 9400 can be rack mounted inside a standard 19 rack It can be mounted using one of two configurations 1 single or 2...

Страница 14: ...9100A 9200A 9400 1 6 Installation Figure 1 1 Dual Rack Mounting Option...

Страница 15: ...s and features except the 9100A is a single channel version the 9200A is a dual channel version and the 9400 has four channels NOTE This manual is common to Models 9100A 9200A and Model 9400 All instr...

Страница 16: ...nnel can be used separately to amplify a unique signal The second mode of operation is the unipolar mode where the signal is applied to one input rectified amplified and output through two separate ou...

Страница 17: ...9100A 9200A 9400 Introduction and Operating Instructions 2 3 Figure 2 1 Model 9100A Single channel High Voltage Power Amplifier Figure 2 2 Model 9200A Dual channel High Voltage Power Amplifier...

Страница 18: ...ear in this manual NOTE A Note symbol contains information relating to the use of this product CAUTION A Caution symbol contains information that should be followed to avoid personal damage to the ins...

Страница 19: ...ctive caps prevent accidental shock hazards Compatible with any of the Tabor arbitrary waveform generators Small case size Options There are no options offered with the model 9400 Manual Changes Manua...

Страница 20: ...he output terminals One can turn on the power to the output terminals only by lifting the safety latch and switching the high voltage on with the HI V switch High voltage on is indicated by a clear li...

Страница 21: ...m 200 V to 200 V with continuous currents up to 50 mA Outputs When used in normal operating mode Unipolar light is off the 9400 can be used as four independent amplifier channels so each of the output...

Страница 22: ...ct the mains cable to the wall socket make sure the AC Voltage Selector switch is set to the correct voltage setting as required for individual countries Information how to install the instrument and...

Страница 23: ...the following The Model 9400 is operating in normal mode when the switch is set to its outside position The instrument is operating in unipolar mode when the switch is pushed in and the front panel Un...

Страница 24: ...ltage path to the output connectors will be blocked when power is applied to the amplifier 2 If you intend to operate the amplifier using its standard bipolar operating mode make sure the rear panel s...

Страница 25: ...ys keep safety distance from the 9400 its output terminals and the high voltage path to the load circuit when high voltage is turned on 10 Note the warning above and flip open the high voltage safety...

Страница 26: ...ly to amplify a unique signal So normal mode of operation provides four separate amplifier channels labeled CH1 through CH4 Figure 2 5 shows an amplified signal using the normal mode of operation CH1...

Страница 27: ...9100A 9200A 9400 Introduction and Operating Instructions 2 13 CH1 Output CH2 Output Figure 2 6 Amplified Unipolar Sine Waveform...

Страница 28: ...9100A 9200A 9400 2 14 Introduction and Operating Instructions This page was intentionally left blank...

Страница 29: ...el Many of the steps covered in this section may expose the individual to potentially lethal voltages that could result in personal injury or death if normal safety precautions are not observed CAUTIO...

Страница 30: ...erve that the High Voltage On light turns off Recommended Test Equipment Recommended test equipment for troubleshooting calibration and performance checking is listed in Table 3 1 below Test instrumen...

Страница 31: ...o potentially lethal voltages that could result in personal injury or death if extreme safety precautions are not observed Do not attempt to perform any of the following tests except if you were train...

Страница 32: ...erator to Channel 1 input Terminate the waveform generator using the 50 feedthrough terminator at the 9400 input Test Procedure 1 Perform Gain Accuracy tests on channel 1 output using Table 3 2 2 Use...

Страница 33: ...channels 2 3 and 4 Table 3 3 Output Bandwidth Large Signals Tests Arb Frequency Oscilloscope Reading Setting Error Limits CH1 CH2 CH3 CH4 Pass Fail 1 kHz 6 Divisions 100 kHz 6 0 5 Divisions 400 kHz 6...

Страница 34: ...ad Resistance x100 high voltage probe 50 feedthrough terminator Preparation 1 Configure the Oscilloscope as follows Time Base 500ns Amplitude 100 V div 2 Connect the 9400 Channel 1 output to the load...

Страница 35: ...vertical trace to show exactly 6 vertical divisions 2 Perform overshoot tests on channel 1 output using Table 3 6 3 Use the same procedure to check channels 2 3 and 4 Table 3 6 Overshoot Tests Paramet...

Страница 36: ...newave distortions should be very similar to those characterized by the main outputs however since the outputs are used for monitoring purpose only some of the characteristics are relaxed Each channel...

Страница 37: ...for the tests 4 Connect the waveform generator to Channel 1 input Terminate the waveform generator using the 50 feedthrough terminator at the 9400 input Test Procedure 1 Using the variable vertical ad...

Страница 38: ...3 10 3 Use the same procedure to check channels 2 3 and 4 Table 3 10 Output Monitor Aberrations Tests Parameter Oscilloscope Reading Tested Error Limits CH1 CH2 CH3 CH4 Pass Fail Rise Time 1 s Fall Ti...

Страница 39: ...k the characteristics of the unipolar circuit Unipolar Mode Indication Equipment Model 9400 Preparation 1 Locate the Unipolar Mode selector switch on the rear panel 2 Locate the Unipolar Mode indicato...

Страница 40: ...and 4 Table 3 12 Unipolar Mode Gain Accuracy Tests Arb Amplitude DMM Reading Setting Error Limits CH1 CH2 CH3 CH4 Pass Fail 2 Vp p 19 30 V 2 V 4 Vp p 38 45 V 4 V 6 Vp p 57 95 V 6 V 8 Vp p 76 80 V 8 V...

Страница 41: ...Performance Checks 3 13 Table 3 13 Unipolar Mode Output Bandwidth Large Signals Tests Arb Frequency Oscilloscope Reading Setting Error Limits CH1 CH2 CH3 CH4 Pass Fail 1 kHz 6 Divisions 100 kHz 6 0 5...

Страница 42: ...ance Checks Do not attempt to calibrate the amplifier before you verify that there is no problem with the functionality of the product A complete set of specification is listed in Appendix A If the in...

Страница 43: ...e inside of the box 2 Identify C105 Using a DMM measure the voltage drop across this capacitor Measurement should verify 12 Vdc 5 3 Lift the front panel high voltage protection latch and turn on the h...

Страница 44: ...ine Bandwidth Adjustment Main Outputs Equipment Waveform Generator Oscilloscope x100 high voltage probe Load Resistance 50 feedthrough terminator Preparation use the same procedure for channels 1 thro...

Страница 45: ...nerator output to the 9400 Channel 1 then 2 3 and 4 input Use 50 feedthrough terminator at the 9400 input Adjustment 1 Identify the trimmers for each channel and make the adjustments below for rise ti...

Страница 46: ...9100A 9200A 9400 Maintenance and Performance Checks 3 18...

Страница 47: ...idth DC to 200 kHz unipolar mode Output Characteristics General Connector BNC Impedance 0 1 Load impedance Resistive recommended for full power bandwidth spec load resistance limited by the output cur...

Страница 48: ...50 VA 120W Weight Approximately 14 lbs 6 5 kg Signal Ground Connected to case ground EMC Certification CE marked Reliability MTBF per MIL HDBK 217E 25 C Ground Benign Safety Designed to meet IEC EN610...

Отзывы: