Calibrating the Autosampler CE
4–12
TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
Figure 4.1
Calibrating a
Cell Lid Position
8. When you are satisfied that the lid is prop-
erly positioned, press the CALIBRATE key
until you see the correct cell lid position on
the display:
Calibrate/Verify
Lid #1 Cell
9. Press ENTER. The screen displays:
Verify (1) or
Store (2) ? 1
LID
IN OPEN
POSITION,
FINGERS
SHOULD BE
EQUIDISTANT
FROM THE
KNOB ON
THE LID.
ALL FOUR
FINGERS
SHOULD
CONTACT
THE LID
AT THE
SAME TIME
WHEN YOU
PRESS THE
GRIP KEY.
GRIPPERS
SIDE VIEW
TOP VIEW
Содержание DSC Autosampler CE
Страница 16: ...TA INSTRUMENTS DSC AUTOSAMPLER CE xvi...
Страница 18: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 2...
Страница 42: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 26...
Страница 44: ...Preparing the Autosampler CE 2 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 50: ...Preparing the Autosampler CE 2 8 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 52: ...Using the Autosampler CE in the Manual Mode 3 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 70: ...Calibrating the Autosampler CE 4 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 106: ...Preparing Samples 5 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 120: ...Preparing Samples 5 16 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 134: ...Diagnostics 6 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 147: ...TA INSTRUMENTS DSC AUTOSAMPLER CE 6 15 Figure 6 5 Turret in Full Up Position Autosampler CE Test Functions...
Страница 190: ...Index I 6 TA INSTRUMENTS DSC AUTOSAMPLER...