6–23
Autosampler CE Test Functions
The screen displays:
Cycle A--B
Enter Point “B”
5. Use the motion keys to put the Autosampler
CE in the second position you wish to cycle
between; then press
ENTER
. The screen
displays:
Cycle A--B
Enter to Start
6. Press
ENTER
. The Autosampler CE
continuously cycles between points A and B.
Be ready to press the HALT key to prevent
possible collisions of the turret arm with the
enclosure, tray, or cell.
Once you are sure that the path the Auto-
sampler CE takes between points A and B
will not result in a collision, you can cycle
this test indefinitely.
7. Press
ESCAPE/CLEAR
to exit the Cycle
AB test.
When you have finished using the Autosampler
CE keypad, remember to close the keypad cover
and finger-tighten the thumbscrews before
starting a method.
t
CAUTION:
NOTE:
Содержание DSC Autosampler CE
Страница 16: ...TA INSTRUMENTS DSC AUTOSAMPLER CE xvi...
Страница 18: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 2...
Страница 42: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 26...
Страница 44: ...Preparing the Autosampler CE 2 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 50: ...Preparing the Autosampler CE 2 8 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 52: ...Using the Autosampler CE in the Manual Mode 3 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 70: ...Calibrating the Autosampler CE 4 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 106: ...Preparing Samples 5 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 120: ...Preparing Samples 5 16 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 134: ...Diagnostics 6 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 147: ...TA INSTRUMENTS DSC AUTOSAMPLER CE 6 15 Figure 6 5 Turret in Full Up Position Autosampler CE Test Functions...
Страница 190: ...Index I 6 TA INSTRUMENTS DSC AUTOSAMPLER...