Calibrating the Autosampler CE
4–22
TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
Figure 4.3
Calibrating a Lid
Storage Position
5. Press the
CALIBRATE
key repeatedly until
you see the correct lid storage position on
the display:
Calibrate/Verify
Lid #1 Storage
6. Press
ENTER
. The screen displays:
Verify (1) or
Store (2) ? 1
7. Select option 2 (Store) to store the calibra-
tion position (press the 2 key, then press
ENTER
).
GRIPPERS
LID
SIDE VIEW
TOP VIEW
IN OPEN POSITION,
FINGERS SHOULD BE
EQUIDISTANT FROM
THE KNOB ON
THE LID.
ALL FOUR FINGERS
SHOULD CONTACT
THE LID AT THE
SAME TIME WHEN
YOU PRESS THE
GRIP KEY.
Содержание DSC Autosampler CE
Страница 16: ...TA INSTRUMENTS DSC AUTOSAMPLER CE xvi...
Страница 18: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 2...
Страница 42: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 26...
Страница 44: ...Preparing the Autosampler CE 2 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 50: ...Preparing the Autosampler CE 2 8 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 52: ...Using the Autosampler CE in the Manual Mode 3 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 70: ...Calibrating the Autosampler CE 4 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 106: ...Preparing Samples 5 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 120: ...Preparing Samples 5 16 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 134: ...Diagnostics 6 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 147: ...TA INSTRUMENTS DSC AUTOSAMPLER CE 6 15 Figure 6 5 Turret in Full Up Position Autosampler CE Test Functions...
Страница 190: ...Index I 6 TA INSTRUMENTS DSC AUTOSAMPLER...