TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
4–23
Performing the Calibration
The screen displays:
Lid #1 Storage
Calibrated
8. Release the lid (press the
RELEASE
key).
Press
RESET SAMPLER
and
ENTER
to
bring the Autosampler CE to the home
position. Press the
CALIBRATE
key until
the correct cell lid position is on the display:
Calibrate/Verify
Lid #1 Storage
9. Press
ENTER
. The screen displays:
Verify (1) or
Store (2) ? 1
10. Press
ENTER
to select default option 1
(Verify).
11. When the turret stops moving, close and
open the grippers again and check your
calibration against the criteria in Figure 4.3.
Remember to release the lid. If the position
is correctly calibrated, go on to step 12. If it
is not, repeat steps 4 through 11 with lid #1.
12. Repeat steps 4 through 11 to calibrate the
storage positions for lids 2 and 3 (the me-
dium-sized and large lids). The screen
displays in steps 5, 7, and 8 are the same
except for the lid numbers, as shown on the
next page.
Содержание DSC Autosampler CE
Страница 16: ...TA INSTRUMENTS DSC AUTOSAMPLER CE xvi...
Страница 18: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 2...
Страница 42: ...Introducing the DSC Autosampler CE TA INSTRUMENTS DSC AUTOSAMPLER CE 1 26...
Страница 44: ...Preparing the Autosampler CE 2 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 50: ...Preparing the Autosampler CE 2 8 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 52: ...Using the Autosampler CE in the Manual Mode 3 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 70: ...Calibrating the Autosampler CE 4 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 106: ...Preparing Samples 5 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 120: ...Preparing Samples 5 16 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 134: ...Diagnostics 6 2 TA INSTRUMENTS DSC AUTOSAMPLER CE...
Страница 147: ...TA INSTRUMENTS DSC AUTOSAMPLER CE 6 15 Figure 6 5 Turret in Full Up Position Autosampler CE Test Functions...
Страница 190: ...Index I 6 TA INSTRUMENTS DSC AUTOSAMPLER...