background image

click

4.  The scrubbed person closes the casing cover from the 

outside without contacting the Power Unit or the inside 
of the casing (Fig. 8–11). The scrubbed person releases 
the “front lid” by pressing on the “Press” button (Fig. 8). 
The scrubbed person places one hand under the cas-
ing cover (Fig. 9) and closes it. They then turn the “front 
lid” 180° until it clicks (Fig. 10).  Always ensure that the 
casing cover and front lid are totally closed before 
 using the handpiece (Fig. 11). 

  

▲ WARNING: 

The scrubbed person must not touch the non-sterile 
Power Unit or the inside of the handpiece in order to 
avoid contamination. Should this happen, the scrubbed 
nurse must be scrubbed again. Should the handpiece be 
contaminated, it must be cleaned and resterilized before 
being used in the operating room.

Removing the Power Unit

1.  The scrubbed person opens the casing as described in 

step 1. The  circulating person removes the Power Unit 
by holding the handle while the scrubbed person holds 
the handpiece firmly (Fig. 12). Ensure that the Power Unit 
does not touch the outer surface of the handpiece to 
avoid contaminating the Power Unit. 
If this occurs follow the  information in the section 
“Cleaning and Disinfection of the Power Unit 
(05.001.602) Directly After the Surgery”. Store the 
Power Unit by using the Adapter for UBC II (05.001.604) 
in the Universal Battery Charger II (05.001.204) when 
not in use. The Universal Battery Charger II should 
 always be turned on (Fig. 13).

  

▲ WARNINGS:

• 

To ensure aseptic conditions, the Power Unit must not 
be removed from the handpiece until the end of 
surgery.

• 

To avoid injuries, always switch OFF the handpiece 
before every manipulation (e.g. inserting/ removing the 
power unit, attachments and/or  cutting tools) and 
before placing it back down.

Figure 9

Figure 8

Figure 11

Figure 10

UNIUM Power Unit
Aseptic Transfer Technique

Figure 12

Figure 13

click

16

Instruction for Use  •  UNIUM

Содержание UNIUM

Страница 1: ...UNIUM POWER TOOL FOR SMALL BONE TRAUMATOLOGY ORTHOPEDIC SURGERY AND STERNOTOMIES Instructions for Use...

Страница 2: ...Attachment and Quick Coupling for Kirschner Wires 20 Burr Attachment 21 Saw Attachments 22 Radiolucent Drive 25 Care and Maintenance General Information 27 Cleaning and Disinfection 29 Preparation Pri...

Страница 3: ...hooting 53 Ordering Information 54 Glossary The following abbreviations are used in the UNIUM Instructions for Use Abbreviation Description W Watt rpm Revolutions per minute Nm Newtonmeter osc min Osc...

Страница 4: ...nst water ingress IPX6 IPX8 IPX9 The handpiece is compatible with a full range of attachments and Small Bone Cutting Tools Introduction UNIUM Portfolio Overview Attachments for UNIUM Modular Handpiece...

Страница 5: ...Handpiece 05 001 612 Size 91 35 29 mm L W H Weight 40 g 0 09 lbs The Top for Sternum is compatible with the UNIUM Reciprocating Saw Handpiece and the following saw blades Reciprocating Saw Blades 511...

Страница 6: ...data is subject to tolerances UNIUM Sterile Cover 05 001 603 To be used for the aseptic transfer technique of the Power Unit detailed on page 15 UNIUM Adapter for UBC II 05 001 604 To charge the Powe...

Страница 7: ...terilization Basket size for UNIUM Reciprocating Saw Handpiece without lid 68 001 651 The insert tray is placed in the basket to hold the attachments during automated washing and sterilization procedu...

Страница 8: ...r the Power Unit Adapter for UBC II and UBC II as described in the Instructions for Use Introduction The UNIUM Portfolio consists of the following items Modular handpiece Reciprocating Saw handpiece P...

Страница 9: ...g always respect the specified Duty Cycles described in this manual For important information regarding electromagnetic compatibility EMC please refer to the corresponding Electromagnetic Compatibilit...

Страница 10: ...se and send it to the DePuy Synthes Service Center Using a damaged device may cause harm to the user or patient If a product falls on the floor fragments may break off This may cause harm to the patie...

Страница 11: ...increase the oscillating drilling speed press down both triggers simultaneously To decrease the speed simultaneously reduce the pressure on both trig gers The oscillating drilling mode helps protect...

Страница 12: ...e slot of the guard The Top for Sternum can be mounted in 4 different positions to accommodate the chosen surgical technique To release the Top for Sternum press both release buttons 4 simultaneously...

Страница 13: ...laced on the underside of the charger indicates which firmware it has If neces sary send the charger to a DePuy Synthes Representative for a firmware update The UBC II includes four independent chargi...

Страница 14: ...he condition of the Power Unit run the self test If the self test has been run and the Service Indicator does not illuminate the Power Unit can be used However if the Power Unit still does not functio...

Страница 15: ...t Fig 1 2 This step is marked on the front lid by an arrow and the words Press and Open Ensure that the casing cover is fully opened Fig 3 2 The scrubbed person places the sterile cover 05 001 603 on...

Страница 16: ...as described in step 1 The circulating person removes the Power Unit by holding the handle while the scrubbed person holds the handpiece firmly Fig 12 Ensure that the Power Unit does not touch the ou...

Страница 17: ...d by the warranty Never use an attachment in reverse mode e g 05 001 254 05 001 257 05 001 258 05 001 259 and 05 001 260 532 015 with a legacy series flexible reamer shaft as this may lead to serious...

Страница 18: ...key Make sure that the teeth of the key engage correctly in the toothed rim of the chuck To remove the cutting tool open the chuck with the key WARNINGS To ensure secure insertion of the cutting tool...

Страница 19: ...ng tool push the coupling sleeve forward and then insert the cutting tool while rotating slightly until it locks in place To disconnect push the coupling sleeve forward and remove the cutting tool The...

Страница 20: ...nt as the torque may not be sufficient AO ASIF Quick Coupling 05 001 251 2 1 Quick Coupling for Kirschner Wires 1 700 rpm Kirschner Wires of any length with a diameter of 0 6 3 2mm can be used with th...

Страница 21: ...the UNLOCK position and remove the burr Insert the new burr as far as possible rotate it slightly until it locks in place and then turn the release sleeve for burrs 1 into the LOCK position until it...

Страница 22: ...aw blade and turn the locking knob 2 clockwise until the saw blade is secured 5 Put the mode switch of the handpiece in the ON or FWD ONLY position Precautions Saw blades labeled Single Use should not...

Страница 23: ...serting and Removing the Saw Blades Pull the sliding sleeve 2 back and then insert the saw blade until it locks into the saw blade coupling 1 Release the sliding sleeve 2 and check that the saw blade...

Страница 24: ...t is set to the unlock position Fig 3 Insert the saw attachment in any position into the attachment coupling of the handpiece until it locks into place To prevent vibrations during operation and to in...

Страница 25: ...slightly Pull back the ring on the attachment to fix the drill Check if the drill bit is seated correctly by gently pulling on it Fig 2 Removing the Drill Bit Follow the same procedure in reverse ord...

Страница 26: ...drill bit are completely in the bone Hitting the medullary nail with the drill bit If overload occurs drilling can be continued after having made the following corrections Correcting the drilling angl...

Страница 27: ...ade of materials such as aluminum titanium and its alloys plastics and compound materials The use of such cleaners should be subject to the data regarding material compatibility in the corresponding d...

Страница 28: ...d equipment materials and personnel in the processing unit Any deviation by the processor from the instructions provided should be properly evaluated for effectiveness and potential adverse consequenc...

Страница 29: ...times using a new clean soft and lint free cloth dampened with a minimum 70 alcohol based disinfectant each time Precautions Do not use solvents to disinfect the Power Unit Sockets must not contact w...

Страница 30: ...se a soft lint free cloth or soft bristled brush to assist in removing gross soil Fig 2 and Fig 3 For cannulations of the handpiece and attachments the cleaning brushes 519 40 519 40 and 532 024 below...

Страница 31: ...nse device with cold tap water for a minimum of 2 minutes Fig 7 Use a syringe or pipette to flush lumens and channels 5 Clean with Detergent Manually clean device under running warm water for a minimu...

Страница 32: ...loth damp ened with a minimum 70 alcohol based disinfectant each time 8 Visually Inspect Device Inspect the device including cannulations sliding sleeves attachment release rings etc for visible soil...

Страница 33: ...e washing basket for the automated cleaning procedure Notes Manual pre cleaning prior to automated cleaning is important to ensure cannulations and other difficult to access areas are clean Alternativ...

Страница 34: ...r Handpiece Basket 68 001 650 with Insert Tray 68 001 654 and Lid 68 001 653 Dimensions Length Width Height Basket without Lid 480 250 100mm Basket with Lid 480 250 108mm Weight Basket 68 001 650 1 46...

Страница 35: ...nside the basket during the automated washing and sterilization procedure available in USA only 532 021 Oscillating Saw Attachment 6 spots for various UNIUM Attach ments 05 001 601 UNIUM Modular Handp...

Страница 36: ...4 F Cleaning agent Rinse 2 Warm deionized water or purified water 40 C 104 F N A Thermal disinfection 5 93 C 199 F N A Dry 40 90 C 194 F N A 4 Inspect Device Remove all devices from the washing basket...

Страница 37: ...th one 1 drop of Synthes Special Oil Fig 1 Once the oil has been ap plied move e g press open close the moving parts sev eral times to properly spread the oil Wipe off excess oil with a cloth This app...

Страница 38: ...nt hole and on the holder of the lever Move the tension lever several times 1 1 1 1 AO ASIF Quick Coupling 05 001 250 05 001 251 Quick Coupling for DHS DCS Triple Reamers 532 015 Quick Coupling for Me...

Страница 39: ...s to be performed by the original manufacturer or an authorized site Do not use excessive oil when lubricating the parts as this may affect the reprocessing Wipe off any excess oil with a cloth Precau...

Страница 40: ...oded devices but return them to the DePuy Synthes Service Center Failing to follow these instructions may lead to damage and malfunction increasing the risk of harm to the user and patient For further...

Страница 41: ...packaging materials Sterile Barrier System e g wraps or re usable rigid container environmental conditions steam quality device materials total mass sterilizer performance and varying cool down time...

Страница 42: ...echnical Service Transporting UNIUM Power Unit The Power Unit is classified as a Li Ion battery contained in equipment according to the International Air Transport Association IATA regulation UN 3481...

Страница 43: ...is 72 hours Do not dismantle open short circuit or shred the Power Unit Precaution Faulty devices must not be reused and should be dis posed of in an environmentally friendly manner and in accordance...

Страница 44: ...cribed If longer periods of constant use are required an additional handpiece and or attachment should be used Depending on the cutting tool used and the load applied the heat generated by the handpie...

Страница 45: ...104 F 10 C 50 F 40 C 104 F 10 C 50 F Relative humidity 95 10 No condensation 95 10 No condensation Atmospheric pressure 1060 hPa 1 06 bar 500 hPa 0 5 bar 1060 hPa 1 06 bar 500 hPa 0 5 bar Altitude 0...

Страница 46: ...manufactured by Gebr Brasseler GmbH Co KG with Synthes acting as the exclusive distributor Handpiece Attachment Cutting Tool Sound Pressure Level LpA in dB A Sound Power Level LwA in dB A Max daily ex...

Страница 47: ...Emission According to EU Directive 2002 44 EC Vibration emissions m s2 tested according to EN ISO 5349 1 Handpiece Attachment Tool Vibration emission m s2 Max daily exposure UNIUM Modular Handpiece 0...

Страница 48: ...cturer s declaration electromagnetic immunity UNIUM Devices are intended for use in the electromagnetic environment of surgery rooms of hospitals except for near active HF surgical equipment The custo...

Страница 49: ...ion distance in meters m Field strengths from xed RF transmitters as determined by an electromagnetic site survey b should be less than the compliance level in each frequency range c Interference may...

Страница 50: ...ut power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power ratin...

Страница 51: ...ront lid should be rotated to open the casing cover Indicates how the sterile cover is placed on top of the handpiece and removed from it during the aseptic transfer Reverse Function is possible with...

Страница 52: ...rradiation Ref 5 2 4 ISO 15223 1 Do not use if package is damaged Ref 5 2 8 ISO 15223 1 Do not re use Ref 5 4 2 ISO 15223 1 Caution United States Federal law restricts this device to sale by or on the...

Страница 53: ...piece or send it to your DePuy Synthes Service Center Oscillating saw attachment vibrates too much The saw blade locking mechanism is not tight Tighten the locking knob of the saw blade coupling Both...

Страница 54: ...I Small Battery Drive II and UNIUM 532 023 Oscillating Saw Attachment II Crescentic Technique for Colibri II Small Battery Drive II and UNIUM 532 026 Large Oscillating Saw Attachment for Colibri II Sm...

Страница 55: ...structions for Use as well as general Instructions for Use are available as PDF files at www e ifu com The trademarks used herein are trademarks of DePuy Synthes www jnjmedicaldevices com en US patent...

Отзывы: