Do not store or transport Power Units haphazardly in a 
box or drawer where they may short-circuit each other or 
be short- circuited by other metal objects. This can dam-
age the Power Unit and generate heat, which can cause 
burns. 

Please note that if the UNIUM Power Unit (05.001.602)   
is damaged, it cannot be transported by aircraft cargo. 
Please refer to the chapter "Repairs and Technical 
 Service" to see when a Power Unit can be considered to 
be damaged. 
Please refer to the chapter "Environmental Conditions"   
to see both transport and storage conditions. 

Servicing 

  

▲ WARNING: 

• 

This device requires regular maintenance  service, at 
least once a year, in order to maintain its safety and 
functionality. This service has to be  performed by the 
original manufacturer or an authorized site.

Warranty and Liability 

The warranty for the tools and accessories does not 
cover normal wear or damage of any kind resulting from 
improper use, improper reprocessing and maintenance, 
damaged seal, use of non-Synthes Cutting Tools and 
 Lubricants or improper storage and transport. The manu-
facturer excludes liability for damage resulting from 
 improper use, neglected or unauthorized maintenance or 
servicing of the tool. In such cases the product will  
no longer be accepted or repaired by DePuy Synthes.
For further information on the servicing, warranty and  
return policy please contact your local DePuy Synthes 
Office.

General Warnings and Precautions

  

▲ WARNINGS: 

• 

For safety reasons, please read the Instructions for Use 
carefully before using the device. 

• 

Do not use any device if the packaging is  damaged or if 
there is any apparent damage to the device. 

• 

Always wear personal protective equipment (PPE) 
including safety goggles when handling with the device. 

• 

To avoid injuries, always switch OFF the handpiece 
before every manipulation (e.g. inserting/removing the 
power unit, attachments and/or cutting tools) and 
before placing it back down. 

• 

Should the handpiece fall on the floor and have visible 
damage, do not use and send it to the DePuy Synthes 
Service Center. Using a damaged device may cause 
harm to the user or patient. 

• 

If a product falls on the floor, fragments may break off. 
This may cause harm to the patient or user due to: 
– loose or sharp fragments. 
– unsterile fragments potentially entering the sterile 
field or coming into contact with the patient.

• 

Synthes Devices are designed and manufactured to 
perform within the scope of their intended use. 
However, if a Power Tool or accessory/attachment 
breaks during use, a visual inspection or a medical 
imaging device (e.g. CT, Radiation De vices, etc.) can 
aid in locating the fragments and/or components of the 
device.

  

▲ Precautions: 

• 

After inserting an attachment, a top for sternum or a 
cutting tool,  always check that it is properly engaged 
by pulling on it gently.

• 

Avoid changing the position of the mode switch while 
the handpiece is operating. Should this  happen, release 
the trigger, turn the mode switch to ON or FWD ONLY 
and press the trigger again to restart the device.

• 

Do not grasp or touch any rotating or moving 
component while the handpiece is operating.

• 

Do not apply excessive pressure with an accessory, 
such as bending or prying. Excessive pressure may 
bend or fracture the accessory and result in tissue 
damage, loss of tactile control, and/or the ejection of 
accessory fragments at a high velocity.

• 

The handpiece must only be operated with a fully 
charged Power Unit. Therefore, ensure that the Power 
Unit is fully charged prior to usage. We recommend 
installing the Power Unit just before use in order to 
prevent unwanted discharge if the handpiece is not in 
the OFF mode. Furthermore, it is recommended that 
the Power Unit is placed into the charger immediately 
after surgery. 

• 

Should the device have corroded parts, discontinue 
use and send it to the DePuy Synthes  Service Center. 

• 

Only use the Universal Battery Charger II (05.001.204) 
to charge the UNIUM Power Unit.  Using a charger that 
does not originate from  Synthes can damage the 
UNIUM Power Unit. 

• 

United States Federal law restricts this device to sale 
by or on order of a physician or other licensed 
healthcare provider.

Introduction
General Information

10

Instruction for Use  •  UNIUM

Содержание UNIUM

Страница 1: ...UNIUM POWER TOOL FOR SMALL BONE TRAUMATOLOGY ORTHOPEDIC SURGERY AND STERNOTOMIES Instructions for Use...

Страница 2: ...Attachment and Quick Coupling for Kirschner Wires 20 Burr Attachment 21 Saw Attachments 22 Radiolucent Drive 25 Care and Maintenance General Information 27 Cleaning and Disinfection 29 Preparation Pri...

Страница 3: ...hooting 53 Ordering Information 54 Glossary The following abbreviations are used in the UNIUM Instructions for Use Abbreviation Description W Watt rpm Revolutions per minute Nm Newtonmeter osc min Osc...

Страница 4: ...nst water ingress IPX6 IPX8 IPX9 The handpiece is compatible with a full range of attachments and Small Bone Cutting Tools Introduction UNIUM Portfolio Overview Attachments for UNIUM Modular Handpiece...

Страница 5: ...Handpiece 05 001 612 Size 91 35 29 mm L W H Weight 40 g 0 09 lbs The Top for Sternum is compatible with the UNIUM Reciprocating Saw Handpiece and the following saw blades Reciprocating Saw Blades 511...

Страница 6: ...data is subject to tolerances UNIUM Sterile Cover 05 001 603 To be used for the aseptic transfer technique of the Power Unit detailed on page 15 UNIUM Adapter for UBC II 05 001 604 To charge the Powe...

Страница 7: ...terilization Basket size for UNIUM Reciprocating Saw Handpiece without lid 68 001 651 The insert tray is placed in the basket to hold the attachments during automated washing and sterilization procedu...

Страница 8: ...r the Power Unit Adapter for UBC II and UBC II as described in the Instructions for Use Introduction The UNIUM Portfolio consists of the following items Modular handpiece Reciprocating Saw handpiece P...

Страница 9: ...g always respect the specified Duty Cycles described in this manual For important information regarding electromagnetic compatibility EMC please refer to the corresponding Electromagnetic Compatibilit...

Страница 10: ...se and send it to the DePuy Synthes Service Center Using a damaged device may cause harm to the user or patient If a product falls on the floor fragments may break off This may cause harm to the patie...

Страница 11: ...increase the oscillating drilling speed press down both triggers simultaneously To decrease the speed simultaneously reduce the pressure on both trig gers The oscillating drilling mode helps protect...

Страница 12: ...e slot of the guard The Top for Sternum can be mounted in 4 different positions to accommodate the chosen surgical technique To release the Top for Sternum press both release buttons 4 simultaneously...

Страница 13: ...laced on the underside of the charger indicates which firmware it has If neces sary send the charger to a DePuy Synthes Representative for a firmware update The UBC II includes four independent chargi...

Страница 14: ...he condition of the Power Unit run the self test If the self test has been run and the Service Indicator does not illuminate the Power Unit can be used However if the Power Unit still does not functio...

Страница 15: ...t Fig 1 2 This step is marked on the front lid by an arrow and the words Press and Open Ensure that the casing cover is fully opened Fig 3 2 The scrubbed person places the sterile cover 05 001 603 on...

Страница 16: ...as described in step 1 The circulating person removes the Power Unit by holding the handle while the scrubbed person holds the handpiece firmly Fig 12 Ensure that the Power Unit does not touch the ou...

Страница 17: ...d by the warranty Never use an attachment in reverse mode e g 05 001 254 05 001 257 05 001 258 05 001 259 and 05 001 260 532 015 with a legacy series flexible reamer shaft as this may lead to serious...

Страница 18: ...key Make sure that the teeth of the key engage correctly in the toothed rim of the chuck To remove the cutting tool open the chuck with the key WARNINGS To ensure secure insertion of the cutting tool...

Страница 19: ...ng tool push the coupling sleeve forward and then insert the cutting tool while rotating slightly until it locks in place To disconnect push the coupling sleeve forward and remove the cutting tool The...

Страница 20: ...nt as the torque may not be sufficient AO ASIF Quick Coupling 05 001 251 2 1 Quick Coupling for Kirschner Wires 1 700 rpm Kirschner Wires of any length with a diameter of 0 6 3 2mm can be used with th...

Страница 21: ...the UNLOCK position and remove the burr Insert the new burr as far as possible rotate it slightly until it locks in place and then turn the release sleeve for burrs 1 into the LOCK position until it...

Страница 22: ...aw blade and turn the locking knob 2 clockwise until the saw blade is secured 5 Put the mode switch of the handpiece in the ON or FWD ONLY position Precautions Saw blades labeled Single Use should not...

Страница 23: ...serting and Removing the Saw Blades Pull the sliding sleeve 2 back and then insert the saw blade until it locks into the saw blade coupling 1 Release the sliding sleeve 2 and check that the saw blade...

Страница 24: ...t is set to the unlock position Fig 3 Insert the saw attachment in any position into the attachment coupling of the handpiece until it locks into place To prevent vibrations during operation and to in...

Страница 25: ...slightly Pull back the ring on the attachment to fix the drill Check if the drill bit is seated correctly by gently pulling on it Fig 2 Removing the Drill Bit Follow the same procedure in reverse ord...

Страница 26: ...drill bit are completely in the bone Hitting the medullary nail with the drill bit If overload occurs drilling can be continued after having made the following corrections Correcting the drilling angl...

Страница 27: ...ade of materials such as aluminum titanium and its alloys plastics and compound materials The use of such cleaners should be subject to the data regarding material compatibility in the corresponding d...

Страница 28: ...d equipment materials and personnel in the processing unit Any deviation by the processor from the instructions provided should be properly evaluated for effectiveness and potential adverse consequenc...

Страница 29: ...times using a new clean soft and lint free cloth dampened with a minimum 70 alcohol based disinfectant each time Precautions Do not use solvents to disinfect the Power Unit Sockets must not contact w...

Страница 30: ...se a soft lint free cloth or soft bristled brush to assist in removing gross soil Fig 2 and Fig 3 For cannulations of the handpiece and attachments the cleaning brushes 519 40 519 40 and 532 024 below...

Страница 31: ...nse device with cold tap water for a minimum of 2 minutes Fig 7 Use a syringe or pipette to flush lumens and channels 5 Clean with Detergent Manually clean device under running warm water for a minimu...

Страница 32: ...loth damp ened with a minimum 70 alcohol based disinfectant each time 8 Visually Inspect Device Inspect the device including cannulations sliding sleeves attachment release rings etc for visible soil...

Страница 33: ...e washing basket for the automated cleaning procedure Notes Manual pre cleaning prior to automated cleaning is important to ensure cannulations and other difficult to access areas are clean Alternativ...

Страница 34: ...r Handpiece Basket 68 001 650 with Insert Tray 68 001 654 and Lid 68 001 653 Dimensions Length Width Height Basket without Lid 480 250 100mm Basket with Lid 480 250 108mm Weight Basket 68 001 650 1 46...

Страница 35: ...nside the basket during the automated washing and sterilization procedure available in USA only 532 021 Oscillating Saw Attachment 6 spots for various UNIUM Attach ments 05 001 601 UNIUM Modular Handp...

Страница 36: ...4 F Cleaning agent Rinse 2 Warm deionized water or purified water 40 C 104 F N A Thermal disinfection 5 93 C 199 F N A Dry 40 90 C 194 F N A 4 Inspect Device Remove all devices from the washing basket...

Страница 37: ...th one 1 drop of Synthes Special Oil Fig 1 Once the oil has been ap plied move e g press open close the moving parts sev eral times to properly spread the oil Wipe off excess oil with a cloth This app...

Страница 38: ...nt hole and on the holder of the lever Move the tension lever several times 1 1 1 1 AO ASIF Quick Coupling 05 001 250 05 001 251 Quick Coupling for DHS DCS Triple Reamers 532 015 Quick Coupling for Me...

Страница 39: ...s to be performed by the original manufacturer or an authorized site Do not use excessive oil when lubricating the parts as this may affect the reprocessing Wipe off any excess oil with a cloth Precau...

Страница 40: ...oded devices but return them to the DePuy Synthes Service Center Failing to follow these instructions may lead to damage and malfunction increasing the risk of harm to the user and patient For further...

Страница 41: ...packaging materials Sterile Barrier System e g wraps or re usable rigid container environmental conditions steam quality device materials total mass sterilizer performance and varying cool down time...

Страница 42: ...echnical Service Transporting UNIUM Power Unit The Power Unit is classified as a Li Ion battery contained in equipment according to the International Air Transport Association IATA regulation UN 3481...

Страница 43: ...is 72 hours Do not dismantle open short circuit or shred the Power Unit Precaution Faulty devices must not be reused and should be dis posed of in an environmentally friendly manner and in accordance...

Страница 44: ...cribed If longer periods of constant use are required an additional handpiece and or attachment should be used Depending on the cutting tool used and the load applied the heat generated by the handpie...

Страница 45: ...104 F 10 C 50 F 40 C 104 F 10 C 50 F Relative humidity 95 10 No condensation 95 10 No condensation Atmospheric pressure 1060 hPa 1 06 bar 500 hPa 0 5 bar 1060 hPa 1 06 bar 500 hPa 0 5 bar Altitude 0...

Страница 46: ...manufactured by Gebr Brasseler GmbH Co KG with Synthes acting as the exclusive distributor Handpiece Attachment Cutting Tool Sound Pressure Level LpA in dB A Sound Power Level LwA in dB A Max daily ex...

Страница 47: ...Emission According to EU Directive 2002 44 EC Vibration emissions m s2 tested according to EN ISO 5349 1 Handpiece Attachment Tool Vibration emission m s2 Max daily exposure UNIUM Modular Handpiece 0...

Страница 48: ...cturer s declaration electromagnetic immunity UNIUM Devices are intended for use in the electromagnetic environment of surgery rooms of hospitals except for near active HF surgical equipment The custo...

Страница 49: ...ion distance in meters m Field strengths from xed RF transmitters as determined by an electromagnetic site survey b should be less than the compliance level in each frequency range c Interference may...

Страница 50: ...ut power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power ratin...

Страница 51: ...ront lid should be rotated to open the casing cover Indicates how the sterile cover is placed on top of the handpiece and removed from it during the aseptic transfer Reverse Function is possible with...

Страница 52: ...rradiation Ref 5 2 4 ISO 15223 1 Do not use if package is damaged Ref 5 2 8 ISO 15223 1 Do not re use Ref 5 4 2 ISO 15223 1 Caution United States Federal law restricts this device to sale by or on the...

Страница 53: ...piece or send it to your DePuy Synthes Service Center Oscillating saw attachment vibrates too much The saw blade locking mechanism is not tight Tighten the locking knob of the saw blade coupling Both...

Страница 54: ...I Small Battery Drive II and UNIUM 532 023 Oscillating Saw Attachment II Crescentic Technique for Colibri II Small Battery Drive II and UNIUM 532 026 Large Oscillating Saw Attachment for Colibri II Sm...

Страница 55: ...structions for Use as well as general Instructions for Use are available as PDF files at www e ifu com The trademarks used herein are trademarks of DePuy Synthes www jnjmedicaldevices com en US patent...

Отзывы: