Distal Locking
1
Insert trocar assembly
Instruments
03.010.063 12.0 mm/8.0 mm Protection Sleeve, 188 mm
03.010.064 8.0 mm/3.2 mm Drill Sleeve, 200 mm
03.010.069 3.2 mm Trocar
Targeted distal locking is used for:
– Ascending screw (level E)
– Distal locking (levels F and G)
Insert the trocar assembly (protection sleeve, drill sleeve and
trocar) through one of the distal holes in the aiming arm.
Make a skin incision and dissect carefully to the bone, to
avoid injury to the surrounding neurovascular structure and
soft tissue.
Advance the trocar assembly and push the trocar down to
the bone. Remove the trocar.
Note:
The holes in the aiming arm for distal locking screws
and the ascending screw are marked with a blue circle.
Technique tip:
Two distinct skin incisions are recommended
when inserting the two distal locking screws.
30
Synthes
MultiLoc Proximal Humeral Nail Technique Guide
Содержание MultiLoc Proximal Humeral Nail