3
Attach centering sleeve for 3.5 mm locking screw
(optional)
Instruments
03.019.019 Centering Sleeve for 3.5 mm Titanium Locking
Screws, and 4.5 mm Titanium MultiLoc Screws
03.019.020 StarDrive Screwdriver, T25, with coupling for
03.019.019, 330 mm
If a 3.5 mm locking screw is used in combination with a
MultiLoc screw, assemble a centering sleeve to the T25
StarDrive screwdriver. Insert the assembly into the protection
sleeve. Turn the screwdriver slightly, until the centering sleeve
aligns with the screw recess. Push the screwdriver to lock the
centering sleeve into the MultiLoc screw.
Remove the screwdriver and protection sleeve.
Notes
– Do not insert a 3.5 mm locking screw through the anterior
screw (level C).
– The centering sleeve may also be inserted manually, see
page 26.
MultiLoc Proximal Humeral Nail Technique Guide
Synthes
25
4
Insert remaining MultiLoc screws
For the remaining screws, repeat Steps 1 through 3.
Carefully inspect the final position of all MultiLoc screws under
imaging in different planes, and ensure that they do not
penetrate the articular surface.
Remove the anterior aiming arm, if used.
Note:
The 3 lateral screws (greater tuberosity, levels A, B,
and D) must be used in any fracture situation as they ensure
the basic stability of the construct.
Содержание MultiLoc Proximal Humeral Nail
Страница 2: ......