33
Implant Removal
Instruments
324.011 LISS Insertion Guide for Distal Femur, left
or
324.012 LISS Insertion Guide for Distal Femur, right
324.043 Fixation Bolt for LISS Insertion Guide
324.022 Drill Sleeve for LISS Insertion Guide
324.044 Stabilization Bolt for LISS Insertion Guide
324.027 Trocar, length 162 mm, for No. 324.022
324.050 Screwdriver Shaft 3.5, hexagonal, length
158 mm
or
324.250 Screwdriver Shaft Stardrive, T25, length
158 mm
324.052 Torque-limiting Screwdriver 3.5, hexagonal
or
314.163 Torque-limiting Screwdriver Stardrive, T25
Remove the implant only after complete consolidation of the
fracture. Remove it in reverse order to the implantation.
First, make the incision for the insertion guide in the path of
the old scar, and mount the insertion guide (see step 1 on
page 16).
Make stab incisions and use the torque-limiting screwdriver
to unlock all screws manually. In a second step, completely
remove all screws with a power tool.