background image

6

Synthes

Universal Battery Charger

Instructions for Use

Putting the Universal charger into operation

Before using the charger, ensure that the mains switch is set
to position «0» (illustration, see page 5). Only use the mains
 cable supplied with the charger to connect it to the mains. 

Use the mains switch to switch on the charger. When the
charging bays are empty and the charger is ready for opera-
tion, this is indicated by all control indicators (red, yellow 
and green) lighting up together. 

Cleaning

Only clean the charger when it is not connected to the
mains. Wipe the Universal Battery Charger occasionally with
a dry cloth (do not use any solvents).

Charging the battery

Put the battery to be charged into a free charging bay the
right way round (see symbols to assist operation, page 4).
Only one battery at a time can be charged in each charging
bay. However, all charging bays can be used simultaneously.
 Following automatic recognition of the battery type and
state, the charging process is started (yellow indicator lights
up). As soon as the battery is completely charged, the
charger switches over to maintenance charging (green in di-
cator lights up). The battery can be removed from the charg-
ing bay at any time, but it is only fully charged when the
green indicator is lit. 

Charging a new battery or one which has not been
used for a prolonged period of time 

New batteries and batteries that have not been used for over
a month and have not been kept stored in the charger will
only reach maximum capacity again after 3 to 5 full charging
cycles. 

Storage of batteries

After each use, remove the battery from the battery casing
and recharge it immediately. Never store used batteries with-
out charging them first. When the batteries are not being
used, always keep them in the charger, which must be left
switched on. Even when the batteries are fully charged
(green indicator lights up), the Synthes Universal Battery
Charger constantly monitors their state and tops up the
charge if necessary. This means that the batteries in the
charger are always ready for use and are stored optimally.
Only store batteries outside the charger if all charging bays
are occupied. 

Do not use batteries that have not been taken directly from
the charger. Otherwise, the charging capac ity may not be
optimal.  

Operating Instructions

Содержание 530.600

Страница 1: ...Instructions for Use Universal Battery Charger...

Страница 2: ......

Страница 3: ...of Product Operating Instructions Troubleshooting 2 4 5 Operating the Universal Battery Charger 6 Cleaning 6 Charging the battery 6 Charging a new battery or one which has not been used for a prolong...

Страница 4: ...Atmospheric 500 1060 hPA 500 1060 hPA pressure 0 5 1 06 bar 0 5 1 06 bar Technical data Operation voltage 100 VAC 240 VAC 50 60 Hz Max current consumption 1 2 A 530 600 2 4 A 530 601 Mains input powe...

Страница 5: ...it switches the drive off automatically when the charge has dropped to the defined residual capacity Only original Synthes products are used these are used properly and the guarantee period has not ex...

Страница 6: ...the battery may not be disposed in the normal waste as it contains Metal Hydride Dispose of or recycle in accordance with lo cal state and federal regulations According to WEEE 2002 96 EC the Universa...

Страница 7: ...bays for Synthes batteries Charge control indicators red yellow green Symbols to assist operation Main switch Fuses 2 5 AT 250 V SPT 5 20 mm Mains cable connector Ventilation openings Description of...

Страница 8: ...mpletely charged the charger switches over to maintenance charging green indi cator lights up The battery can be removed from the charg ing bay at any time but it is only fully charged when the green...

Страница 9: ...down and is automatically resumed after a maximum of 25 minutes Red indicator lights up The battery is defective and must be replaced No indicator lights up on charger The charger has no mains power o...

Страница 10: ...charger on again 4 Have charger checked and if necessary repaired by the official Synthes manufacturer or an authorised sales outlet Have charger checked and if necessary repaired by the official Synt...

Страница 11: ...l down in the charging bay After a maximum of 45 minutes the yellow indicator also lights up After a further 25 minutes the charging process is automatically resumed If the yellow indicator does not l...

Страница 12: ...in charger or battery Replace battery Replace battery Replace battery Replace battery Have charger and battery checked and if necessary repaired by the official Synthes manufacturer or an authorised...