background image

Synthes

7

Charge control indicators

There is a red, a yellow and a green illuminated indicator for
each charging bay. These have the following meanings
 (illustration, see page 5): 

All indicators light up:

The charger is ready for operation. 

Yellow indicator lights up:

The battery is being charged. 

Green indicator lights up:

The battery is ready for use. For optimum storage
and charge maintenance it should be left in the
charger until used again. 

Only red indicator flashes: 

The battery is too warm and it has to cool down be-
fore charging is automatically resumed; or the
charger was not able to detect the cell type. If the
battery is not noticeably warm and the indicator
continues to flash on and off for more than 45 min-
utes, remove the battery from the charger briefly
and then re-insert it. If the indicator continues to
flash, the battery is defective and must be replaced. 

Red indicator flashes 
and yellow indicator lights up: 

The battery is too warm. Charging is interrupted to
allow it to cool down and is automatically resumed
after a maximum of 25 minutes. 

Red indicator lights up: 

The battery is defective and must be replaced.

No indicator lights up on charger:

The charger has no mains power or is defective. If it
is defective, it must be checked and, if necessary,
 repaired by the  official Synthes manufacturer or an
authorised sales outlet. 

No indicator lights up on one charging bay: 

If there is a battery inserted, remove it, switch the
charger off briefly and then switch it on again. If all
three indicators light up, the battery is defective. If
none of the indicators light up again, then the
charging bay is defective and the charger must be
checked and, if necessary, repaired by the official
Synthes manufacturer or an authorised sales outlet.

Caution

– After a power cut or a changeover to emergency power,

the charger is automatically switched on again. 

– When charging batteries other than those specified on

page 3 there is a risk of explosion. 

– Batteries that are no longer usable must be disposed of as

hazardous waste. Please observe local, state and federal
regulations.

– Further information on the storage and use of Synthes

 batteries can be found in the respective instruction leaflets. 

Содержание 530.600

Страница 1: ...Instructions for Use Universal Battery Charger...

Страница 2: ......

Страница 3: ...of Product Operating Instructions Troubleshooting 2 4 5 Operating the Universal Battery Charger 6 Cleaning 6 Charging the battery 6 Charging a new battery or one which has not been used for a prolong...

Страница 4: ...Atmospheric 500 1060 hPA 500 1060 hPA pressure 0 5 1 06 bar 0 5 1 06 bar Technical data Operation voltage 100 VAC 240 VAC 50 60 Hz Max current consumption 1 2 A 530 600 2 4 A 530 601 Mains input powe...

Страница 5: ...it switches the drive off automatically when the charge has dropped to the defined residual capacity Only original Synthes products are used these are used properly and the guarantee period has not ex...

Страница 6: ...the battery may not be disposed in the normal waste as it contains Metal Hydride Dispose of or recycle in accordance with lo cal state and federal regulations According to WEEE 2002 96 EC the Universa...

Страница 7: ...bays for Synthes batteries Charge control indicators red yellow green Symbols to assist operation Main switch Fuses 2 5 AT 250 V SPT 5 20 mm Mains cable connector Ventilation openings Description of...

Страница 8: ...mpletely charged the charger switches over to maintenance charging green indi cator lights up The battery can be removed from the charg ing bay at any time but it is only fully charged when the green...

Страница 9: ...down and is automatically resumed after a maximum of 25 minutes Red indicator lights up The battery is defective and must be replaced No indicator lights up on charger The charger has no mains power o...

Страница 10: ...charger on again 4 Have charger checked and if necessary repaired by the official Synthes manufacturer or an authorised sales outlet Have charger checked and if necessary repaired by the official Synt...

Страница 11: ...l down in the charging bay After a maximum of 45 minutes the yellow indicator also lights up After a further 25 minutes the charging process is automatically resumed If the yellow indicator does not l...

Страница 12: ...in charger or battery Replace battery Replace battery Replace battery Replace battery Have charger and battery checked and if necessary repaired by the official Synthes manufacturer or an authorised...

Страница 13: ......

Страница 14: ......

Страница 15: ......

Страница 16: ...0123 036 000 725 SE_057689 AB 41080008 Synthes 2008 Synthes Inc or its affiliates All rights reserved Synthes is a trademark of Synthes Inc or its affiliates Presented by...

Отзывы: