Red indicator flashes. 

Red indicator flashes and yellow indi ca-
tor lights up. 

Battery does not have sufficient power. 

Problem

Possible causes

Remedy

Battery has overheated. 

Charger has been unable to detect
 battery type correctly. 

Battery has overheated and charging
process has been interrupted to allow it
to cool down.

Battery has been stored outside charger
and not been used for over 
a month.

Over-discharged battery.

Battery not ready for operation. 

Battery’s expected expiry date has been
reached. 

Let battery cool down in the charging
bay. After a maximum of 45 minutes
the yellow indicator also lights up.
 After a further 25 minutes the charging
process is automatically resumed. 

If the yellow indicator does not light up
after a maximum of 45 minutes, 
the battery is defective and must be
 replaced. 

Charging process is automatically 
resumed after a maximum of 25 min-
utes. 

3 to 5 charging/discharging cycles 
are necessary for the battery 
to reach maximum capacity again. 

Completely discharged battery was not
recharged immediately after use and
has not been used for several weeks.
3 to 5 charging/discharging cycles are
necessary for battery to reach maxi-
mum capacity again.

Charge battery until green indicator
lights up. 

Replace battery. 

Содержание 530.600

Страница 1: ...Instructions for Use Universal Battery Charger...

Страница 2: ......

Страница 3: ...of Product Operating Instructions Troubleshooting 2 4 5 Operating the Universal Battery Charger 6 Cleaning 6 Charging the battery 6 Charging a new battery or one which has not been used for a prolong...

Страница 4: ...Atmospheric 500 1060 hPA 500 1060 hPA pressure 0 5 1 06 bar 0 5 1 06 bar Technical data Operation voltage 100 VAC 240 VAC 50 60 Hz Max current consumption 1 2 A 530 600 2 4 A 530 601 Mains input powe...

Страница 5: ...it switches the drive off automatically when the charge has dropped to the defined residual capacity Only original Synthes products are used these are used properly and the guarantee period has not ex...

Страница 6: ...the battery may not be disposed in the normal waste as it contains Metal Hydride Dispose of or recycle in accordance with lo cal state and federal regulations According to WEEE 2002 96 EC the Universa...

Страница 7: ...bays for Synthes batteries Charge control indicators red yellow green Symbols to assist operation Main switch Fuses 2 5 AT 250 V SPT 5 20 mm Mains cable connector Ventilation openings Description of...

Страница 8: ...mpletely charged the charger switches over to maintenance charging green indi cator lights up The battery can be removed from the charg ing bay at any time but it is only fully charged when the green...

Страница 9: ...down and is automatically resumed after a maximum of 25 minutes Red indicator lights up The battery is defective and must be replaced No indicator lights up on charger The charger has no mains power o...

Страница 10: ...charger on again 4 Have charger checked and if necessary repaired by the official Synthes manufacturer or an authorised sales outlet Have charger checked and if necessary repaired by the official Synt...

Страница 11: ...l down in the charging bay After a maximum of 45 minutes the yellow indicator also lights up After a further 25 minutes the charging process is automatically resumed If the yellow indicator does not l...

Страница 12: ...in charger or battery Replace battery Replace battery Replace battery Replace battery Have charger and battery checked and if necessary repaired by the official Synthes manufacturer or an authorised...

Страница 13: ......

Страница 14: ......

Страница 15: ......

Страница 16: ...0123 036 000 725 SE_057689 AB 41080008 Synthes 2008 Synthes Inc or its affiliates All rights reserved Synthes is a trademark of Synthes Inc or its affiliates Presented by...

Отзывы: