4036B Operations and Maintenance Manual Rev. E 

11 

 

  

2.3 Cleaning 

 

                   

Warning

 

Do not spray or use too much liquid when cleaning the unit. Liquid can enter the unit and damage 
sensitive electronic components. 

 

Clean the main chassis with a soft cloth dampened with a mild soap and water solution.

 

2.4 Installing the 4036B    

 

The 4036B is shipped ready for installation into a standard 19" (48.3 cm) rack. Provisions are 
included for mounting General Devices slides with part numbers C-300-S-126, -128 and -130. 

2.4.1 Required materials for installation 

North American or European IEC power cord. One or the other will be supplied with the unit.  

Customer supplied double shielded RG223 cables with BNC connectors from source, and to 

next devices in system.  

Customer supplied shielded LAN cable for network connection (RJ-45). 

Rack-mount slide kit from General Devices, C-300-S-126, -128 or -130 (Optional) 

Rack mounting screws.  

Screwdriver for the rack mount screws, and slide as needed.  

2.4.2 Making Connections 

2.4.2.1 Input Power 

AC INPUT POWER - The input power to the unit is supplied through a detachable 3-prong 
power cable. First plug the female end into the male IEC-320 plug on the rear of the unit, 
then plug the male end of the cable into a 100V~ to 240V~, 50/60 Hz power source. 

Warning  

Ensure that this power supply cord is connected to a properly grounded mains receptacle to 
ensure safety. 

Since the unit does not have a AC Mains Power Switch, both the Appliance Inlet Connector 
and the Plug on the detachable type power supply cord are considered to be suitable 
Disconnect Means for disconnecting the unit from the AC Mains Supply. If the rear of the 
unit is not accessible after installation in the instrument rack, the User is responsible for the 
provision of a suitable external AC Disconnect Means for the unit. 

2.4.2.2 Input Signal 

Connect the input signal cable to the 

INPUT 

BNC connector on the back of the unit. 

 

 

 

Содержание TSC 4036B

Страница 1: ...TSC 4036B 1 x 15 RF Distribution Amplifier Operations and Maintenance Manual Revision E November 2010 ...

Страница 2: ...2 4036B Operations and Maintenance Manual Rev E Contact Information www symmetricom com support symmetricom com Phone Worldwide Main Number 1 408 428 7907 USA toll free 1 888 367 7966 1 888 FOR SYMM ...

Страница 3: ...Revision History Revision Description Date Approved A Initial Release 12 23 03 GAR B Added CE information 1 13 04 GAR C Added RF gain setting information 1 29 04 GAR D Changed pollution note and wording for AC disconnect Added requirement for double shielded BNC cables Added call out for rear panel Gnd Stud 2 11 04 GAR E Revised warranty shipping and Declaration of Conformity information Changed p...

Страница 4: ...he gain 13 3 Monitoring the 4036B 14 3 1 Accessing the System 14 3 2 Checking System Status 14 3 2 1 Checking Alarms and Input Frequency 14 3 2 2 Checking Model Number and Software Version 15 3 3 Understanding Alarm Output 15 4 Troubleshooting the 4036B 16 4 1 Replacing Fuses 16 4 2 Verifying Operational Problems 16 5 Shipping Information 18 5 1 Warranty Information 18 5 2 Shipping Information 18 ...

Страница 5: ...t the manual as well as on the unit itself This symbol means the following information is a note that gives you important information that may affect how you use the Caution refer to manual Read all instructions in Manual before using this product CE marking attesting compliance to applicable European Directives Caution Risk of Electrical shock Fuse symbol Mains Power is ON Mains Power is OFF 5 Th...

Страница 6: ...ort NETWORK DO NOT CONNECT TO TELECOM CONNECTIONS THAT CARRY HAZARDOUS VOLTAGES 101010 Serial port connection DO NOT CONNECT TO TELECOM CONNECTIONS THAT CARRY HAZARDOUS VOLTAGES Earth terminal symbol Used to indicate an earth ground connection to chassis ...

Страница 7: ...6 describes how to replace fuses and verify operational problems Chapter 5 Warranty and Shipping Information on page 18 explains how to contact Symmetricom for warranty service and provides shipping guidelines Chapter 6 Declaration of Conformity on page 20 provides the information pertaining to CE marking of this product Appendix A Specifications on page 22 contains the detailed specifications for...

Страница 8: ...he front panel provides green LED status for the input and green red for all output signals as well as indicating an input of 1 MHz 5 MHz or 10 MHz A green LED on the front panel also provides power status Figure 1 shows the 4036B s front panel and Figure 2 shows the 4036B s rear panel Figure 1 Front panel Figure 2 Rear panel POWER Green when power is on INPUT FREQUENCY Lit LED shows whether input...

Страница 9: ...4036B Operations and Maintenance Manual Rev E 9 This page intentionally left blank ...

Страница 10: ...back Verify that input line voltage and current capacity are within specifications before turning on power to the unit Disconnect all sources of input power before removing the top cover of this unit Operating and maintenance personnel must receive proper training before installing or maintaining electrical equipment 2 2 Unpacking To unpack the TSC 4036B unit 1 Unpack and carefully inspect the uni...

Страница 11: ...0 S 126 128 or 130 Optional Rack mounting screws Screwdriver for the rack mount screws and slide as needed 2 4 2 Making Connections 2 4 2 1 Input Power AC INPUT POWER The input power to the unit is supplied through a detachable 3 prong power cable First plug the female end into the male IEC 320 plug on the rear of the unit then plug the male end of the cable into a 100V to 240V 50 60 Hz power sour...

Страница 12: ...IP address For more detailed information see the Xport User Manual Section 3 3 discusses several different ways that you can assign IP addresses Chapter 4 explains how to configure a static IP address You can download the Xport User Manual from the Lantronix Web site as an Adobe Acrobat PDF file Go to http www lantronix com products eds xport XPort_UM_900 270 pdf If you want to configure a static ...

Страница 13: ...on 0 3 Set the IP address and follow on screen commands to save the setting Sending an I command will turn on the three front panel input LEDs for 2 seconds before returning to normal operation This provides the capability for the user to ensure they are communicating over the Ethernet connection to the correct device via front panel feedback If you move the 4036B to a different network hub after ...

Страница 14: ...current alarms latched alarms and the type of input signal To check system status Type S case sensitive You do not need to type a carriage return or line feed The system returns Scccc llll x r n where cccc is current alarms llll is latched alarms since the last status request Both cccc and llll are hexadecimal numbers with each binary bit position representing one output The MSB is the input signa...

Страница 15: ...ystem returns Vxx r n where xx is the software version Example V80 Sending an I command will turn on the three front panel input LEDs for 2 seconds before returning to normal operation This provides the capability for the user to ensure they are communicating over the Ethernet connection to the correct device via front panel feedback 3 3 Understanding Alarm Output The 4036B automatically publishes...

Страница 16: ...lat head screwdriver Replacement fuse for a standard IEC 320 power entry module with fuse 5 x 20 mm 1 A 250 V Time lag fuse To replace a fuse 1 Disconnect the power cable from the back of the 4036B 2 Using a small screwdriver open the fuse cover on the back of the 4036B 3 Replace the old fuses as necessary 4 Close the fuse cover 5 Reconnect the power cable to the back of the 4036B 4 2 Verifying Op...

Страница 17: ...4036B Operations and Maintenance Manual Rev E 17 This page intentionally left blank ...

Страница 18: ...er 1 408 428 7907 USA toll free 1 888 367 7966 1 888 FOR SYMM 5 2 Shipping Information If you need to ship this system for any reason including returning equipment to Symmetricom for warranty service follow these shipping instructions Failure to follow these instructions may damage your system 5 2 1 Packing Instructions Always ship the 4036B appropriately packaged to protect it from damage prefera...

Страница 19: ...4036B Operations and Maintenance Manual Rev E 19 This page intentionally left blank ...

Страница 20: ...MODEL NO TSC 4130A CONFORMS TO THE FOLLOWING EUROPEAN UNION DIRECTIVES Safety 2006 95 EC Low Voltage Directive LVD IEC 61010 1 2001 EN 61010 1 2001 Electromagnetic Compatibility 2004 108 EC Electromagnetic Compatibility EMC Directive EN61326 2006 EMC Requirements for Measurement Control and Laboratory Equipment Class B EN61000 3 2 2006 Harmonic Current Emissions EN61000 3 3 1995 w A1 2001 A2 2006 ...

Страница 21: ...ve 2002 95 EC This product falls under the category of Monitoring and Control Instruments Equipment Category 9 as defined in Annex 1A of the WEEE 2002 96 EC Directive which is excluded from the RoHS Directive 2002 95 EC reference Article 2 paragraph 1 requirements We declare that the equipment specified above conforms to the above Directives and Standards 03 November 2010Robert MengelbergComplianc...

Страница 22: ... reflects the overall Power Supply ratings For UL and CE compliance the Power Supply must only be operated at 100 240 VAC 50 60 Hz Power Inlet Type IEC 60320 sheet C14 Power Supply Cord Set 18 AWG 0 75 mm2 minimum Power Mains Fuse 2 250V 1A Time lag 5x20 mm Signal Input and output Frequency 1 20 MHz Impedance 50 Ω 5 Ω Input level 0 1 V rms 13 dBm Connectors Input 1 BNC Output 15 BNC LAN RJ 45 Gain...

Страница 23: ...4036B Operations and Maintenance Manual Rev E 23 Item Specification SSB phase noise 1 Hz 135 dBc 10 Hz 145 dBc 100 Hz 155 dBc 1 kHz 163 dBc 10 kHz 163 dBc ...

Страница 24: ...table for continuous operation Table 4 lists the environmental specifications for the 4036B Table 4 Environment specifications Item Temperature Relative Humidity Altitude In Use 0 C to 50 C 0 to 95 non condensing 3 000 meters 9 843 feet Storage 40 C to 70 C 0 to 95 non condensing Transportation 40 C to 70 C 0 to 95 non condensing A 3 Physical Specifications Table 5 lists the physical specification...

Страница 25: ...riage return CNR Command and Response DHCP Dynamic Host Configuration Protocol ESD electrostatic discharge LED light emitting diode LSB least significant bit MSB most significant bit PDF portable document format PWA printed wiring assembly RF radio frequency Glossary ...

Страница 26: ...put signal 8 conventions typographic 7 DeviceInstaller software 12 13 Ethernet connecting 14 port 8 12 14 fuse replacing 16 symbol 5 gateway address assigning 12 information shipping 18 warranty 18 20 input connector 8 LEDs 8 input frequency checking 14 input plug symbol 5 input power connecting 11 installing a 4036B 11 instructions packing 18 safety 10 IP address assigning 12 LAN port symbol 5 La...

Страница 27: ...ion checking 15 specifications electrical 22 environment 23 physical 23 subnet mask assigning 12 symbols 5 CE marking 5 fuse 5 input plug 5 LAN port 5 system commands 14 system status checking 14 troubleshooting procedures 16 typographic conventions 7 unpacking a 4036B 10 warning 5 symbol 5 warranty information 18 20 Xport Ethernet to RS 232 converter 12 user manual 12 ...

Отзывы: