4. Calibration
With and select according to your
significance, with leave the menu
without any change, with switch to
a deeper menu level.
4.1 Calibration Factor
Global calibration factor of the measuring
on the display and as well the output range
from 0.01 to 9.99 - Setting is 1.0
With set the value to 0.0 to start enter the
numbers of the measuring range, with
transfer the entry and leave the menu level.
4.3 Number of
Calibration Points
8
Calibration
7 8 9
Cal.-Factor
4 5 6
1.0
1 2 3
C
0
8
8
C
C
Calibration
7 8 9
Segment Points
4 5 6
2
1 2 3
C
0
8
C
8
£
¤
4. Calibration
£
4.1 Cal. Factor
1.0
¤
4.2 Filter
0.1 s
4.3 Aux. Points 2
C
4.4 Calibration
8
1.0
Set the number of the auxiliary points:
2 ... 5 points.
With set the value to 0.0 to start enter the
numbers of the measuring range, with
transfer the entry and leave the menu level.
4.4 Calibration
With and select according to your
significance, with leave the menu
without any change, with transfer the
entry and switch to a deeper menu level.
C
8
£
¤
4.4 Calibration
£
4.4.1 Val. P1
1. 0.00
¤
4.4.2 Calibration P1
4.4.1 Val P2 1.58
C
4.4.2 Calibration P2
8
q
The menu is started by an invisible key in
the bottom left hand
corner of
the
touch-screen-panel. Now press approx. 5 seconds until the menu
appears.
If the temperature indication is activated, the button for the
temperature indication is in the upper right corner, in this case must be changed
into the temperature indication first, in order to be able to access the menu.
It is sufficient to carry out a two-point-calibration (normally min and max).
operation
. (This value can be adjusted later on.)
Thus the basic function of the measuring system is given and it is now ready for
operation.
4.4.1 Calibration Point 1 - Measuring Value
Measuring value in phys. units.
Range: Measuring start - Measuring end
With set the value to 0.0 to start enter the
numbers of the measuring range, with
transfer the entry and leave the menu level.
4.4.2 Calibration Point 1 - Raw Value
Indicate the initial value to the value
displayed, if pressed . With leave the
menu without any change.
8
C
8
C
Calibration
7 8 9
Calibration Point 1
4 5 6
0.0
kg/h
1 2 3
C
0
8
Calibration
Calibration Point 1
0.015663
C
Akt.: 0.015009
8
All other points are calibrated as the first one.
Min-Point
Set Calibration
Set point 1 to 0
(empty)
, w
ith
no
material flow but system
running
.
Max-Point
Set Calibration
Set point 2 to maximum flow rate with normal conveying and