Operating Instructions
27
Connecting Diagram for C3-Box
Programming
In addition to the basic SolidFlow Start-up with one sensor it is necessary to switch on the sensors which should
be used for the measurement (normally all sensors which are installed). As well every sensor has his own
calibration-factor which can be used for adjusting the influence which each sensor gives to the measurement
result (normally factor 1.0 for all sensors).
This leads to the following menu-structure for the sensor calibration:
4.4 Calibration Sensor 1
4.4.1 Sensor on / Off
4.4.2 Calibration Factor Sensor
4.4.3 Calibration Point 1 Value
4.4.4 Calibration Point 1 Raw Value
4.4.5 Calibration Point 2 Value
4.4.6 Calibration Point 2 Raw Value
analogue for all further calibration points
4.5 Calibration Sensor 2
(like sensor 1)
4.6 Calibration Sensor 3
(like sensor 1)
The calibration of each sensor is to be done as it is described for the FME 100.
NOTE:
It is important to program the total flow as a pre-set value for each sensor. The FME 300 is
calculating the average value of all single measurements automatically and this average value
will be given to the analog output.
Sensor 1
Sensor 2
Sensor 3
Evaluation Unit