AMI INSPECTOR Oxygen
Program List and Explanations
A-96.250.701 / 140716
65
5.3.2.342
Calendar
:
5.3.2.342.1
Start time
: The programmed start time is valid for each of the
programmed days. To set the start time see
5.3.2.341, p. 64
.
Range: 00:00:00–23:59:59
5.3.2.342.2
Monday
: Possible settings, on or off
to
5.3.2.342.8
Sunday
: Possible settings, on or off
5.3.2.44
Run Time
: see Interval
5.3.2.54
Delay
: see Interval
5.3.2.6
Signal Outputs
: see Interval
5.3.2.7
Output/Control
: see Interval
The relay will be switched via the Profibus input. No further
parameters are needed.
5.3.4
Input:
The functions of the relays and signal outputs can be
defined depending on the position of the input contact, i.e. no
function, closed or open.
5.3.4.1
Active:
Define when the input should be active:
5.3.4.2
Signal Outputs:
Select the operation mode of the signal outputs
when the relay is active:
5.3.2.1
Function = Fieldbus
No:
Input is never active.
When closed
Input is active if the input relay is closed
When open:
Input is active if the input relay is open
Cont.:
Signal outputs continue to issue the measured
value.
Hold:
Signal outputs issue the last valid measured
value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off:
Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.