56
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Program List and Explanations
5.2.1.40
Scaling:
Enter beginning and end point (Range low & high) of the
linear or logarithmic scale. In addition, the midpoint for the bilinear
scale.
Parameter: Oxygen.
Range low: 0.00 ppb –20.00 ppm
Range high: 0.00 ppb –20.00 ppm
Parameter: Temperature
Range low: -30 to + 130 °C
Range high: 30 to + 130 °C
Parameter: Sample flow
Range low: 0–50 l/h
Range high: 0–50 l/h
Parameter: Saturation
Range low: 0–200 %
Range high: 0–200 %
X
Measured value (logarithmic)
20
1
0
1
2
3
4
10
100
1’000
10’000
10 12
(0 - 20 [mA])
0 / 4
4
2
6
(4 - 20 [mA])
[mA]
X