36
A-96.250.701 / 140716
AMI INSPECTOR Oxygen
Maintenance
6.6.1
Activate SWAN Quality assurance procedure
Enable quality assurance procedure on the process monitor(s)
which shall be checked by selecting the quality level in menu
5.1.2.1. The corresponding submenus are then activated.
NOTICE:
The activation is necessary the first time only.
6.6.2
Pre-test
Reference instrument: AMI Inspector Oxygen:
– Check certificate; reference instrument certificate not older
than one year.
– Check battery; Battery of the AMI Inspector Oxygen should
be completely charged. Remaining operating time on
display minimum 20 hours.
– Make sure that all instruments are set to the same
temperature compensation.
– Sensor is in working condition
On-line instrument: Monitor AMI Oxytrace:
– Good order and condition; Flow cell free of particles,
Sensor surface free of deposits.
– Check message list; Review the message list in menu 1.3
and check for frequently occurring alarms (as for example
flow alarms). If alarms occur frequently remove cause
before starting the procedure.
6.6.3
Connect the sample lines
See corresponding chapter in the manual of the process monitor
which shall be checked with a reference instrument.
The choice of sampling depends strongly on local conditions on
site. Possible sampling:
via sample point,
via T-fitting or
via piggyback/downstream
NOTICE:
•
avoid ingress of air, use screwed fitting,
•
sample as near as possible to the process monitor,
•
wait approx. 10 minutes, whilst measurement is running,
until measurement value and temperature are stabilized.