SV 258 PRO User Manual
39
Note: To ensure saving of Summary results in a file
you should switch on the Auto Save function in
SVAN 958A (path: <Menu> / Files / Save Options /
Auto Save: Number). In this case all files are
automatically named in accordance with the rule: same
prefix (string of letters) and number (string of digits),
which will be increased by one each time the new file
is created.
Note: The Auto Save function is available only for the Integration period not lower than 25
seconds!
Note: Time history files are created automatically if Time history is On with predefined name
Buffer_xx, where xx is a number, which will be increased by one each time the new file is
created.
In the
EVENT TRIGGER
tab, you can:
1. configure events for triggering markers, audio recording and alarms in the
Alarm triggers
section,
2. enable trigger of time-history logging, set logger pre and posttrigger (this section appears if
Time history
is
On
in the
STORAGE
tab) and
3. configure notification in the
SMS & e-mail
section.
Events are specified as a combination of superimposed
CONDITIONS
(logical AND) such as specific time
intervals (
TIME PERIODS
) in which measurement threshold levels are exceed or system events occur
(
TRIGGER
).
Each Event may be connected with special triggers
(
Actions
) such as: time-history data recording to the
logger file (
Logger
), generation of the alarm signal on the I/O socket (
Alarm flag
), alarm SMS sending
(
SMS alarm
) or alarm e-mail sending (
E-mail alarm
).
1
2
3