LAMBDA DG-4 & DG-5 OPERATION MANUAL – REV.
3.05C
(
20150414
)
38
Table 4-4. DG-5-Only filter position movement commands.
Move to
Nominal
Filter
Position
Value
(Decimal;
hexadecimal; binary)
Move on
…
Atte-
nuation
Physical
Filter
Position
Notes
0
0, 00, 00000000
Command
0% 0
Light is blocked (“Shutter Closed”)
1
1; 01; 00000001
Command 100% 1
1
st
filter (18mm) at full
2
2; 02; 00000010
Command 100% 2
2
nd
filter (18mm) at full
3
3; 03; 00000011
Command 100% 3
3
rd
filter (18mm) at full
4
4; 04; 00000100
Command 100% 4
4
th
filter (25mm) at full
5
5; 05; 00000101
Command 100% 5
5
th
filter (25mm) at full
6
6; 06; 00000110
Command
50% 1
1
st
filter (18mm) at half
7
7; 07; 00000111
Command
50% 2
2
nd
filter (18mm) at half
8
8; 08; 00001000
Command
50% 3
3
rd
filter (18mm) at half
9
9; 09; 00001001
Command 50% 4
4
th
filter (25mm) at half
10
10; 0A; 00001010
Command
50% 5
5
th
filter (25mm) at half
11
11; 0B; 00001011
Command
33% 1
1
st
filter (18mm) at 1/3
12
12; 0C; 00001100
Command
33% 2
2
nd
filter (18mm) at 1/3
13
13; 0D; 00001101
Command 33% 3
3
rd
filter (18mm) at 1/3
14
14; 0E; 00001110
Command
33% 4
4
th
filter (25mm) at 1/3
15
15; 0F; 00001111
Command
33% 5
5
th
filter (25mm) at 1/3
16
16; 10; 00010000
Trigger 0% 0 Light is blocked (“Shutter Closed”)
17
17; 11; 00010001
Trigger 100% 1 1
st
filter (18mm) at full
18
18; 12; 00010010
Trigger 100% 2 2
nd
filter (18mm) at full
19
19; 13; 00010011
Trigger 100% 3 3
rd
filter (18mm) at full
20
20; 14; 00010100
Trigger 100% 4 4
th
filter (25mm) at full
21
21; 15; 00010101
Trigger 100% 5 5
th
filter (25mm) at full
22
22; 16; 00010110
Trigger 50% 1 1
st
filter (18mm) at half
23
23; 17; 00010111
Trigger 50% 2 2
nd
filter (18mm) at half
24
24; 18; 00011000
Trigger 50% 3 3
rd
filter (18mm) at half
25
25; 19; 00011001
Trigger 50% 4 4
th
filter (25mm) at half
26
26; 1A; 00011010
Trigger 50% 5 5
th
filter (25mm) at half
27
27; 1B; 00011011
Trigger 33% 1 1
st
filter (18mm) at 1/3
28
28; 1C; 00011100
Trigger 33% 2 2
nd
filter (18mm) at 1/3
29
29; 1D; 00011101
Trigger 33% 3 3
rd
filter (18mm) at 1/3
30
30; 1E; 00011110
Trigger 33% 4 4
th
filter (25mm) at 1/3
31
31; 1F; 00011111
Trigger 33% 5 5
th
filter (25mm) at 1/3
NOTE: “Move on Command” means the light beam is moved to the specified position
immediately upon receipt of the external command. “Move on Trigger” means move the light
beam only until triggered via Strobe, Synch, or Synch gated to Strobe.
Содержание Lambda DG-4
Страница 3: ......
Страница 4: ......
Страница 9: ......
Страница 14: ......
Страница 18: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 4 This page intentionally blank ...
Страница 26: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 12 This page intentionally blank ...
Страница 48: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 34 This page intentionally blank ...
Страница 63: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 49 ...
Страница 64: ......
Страница 70: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 56 This page intentionally blank ...
Страница 72: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 58 This page intentionally blank ...
Страница 82: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 68 This page intentionally blank ...
Страница 86: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 72 NOTES ...