Chapter 3: Setting Up the Monitor
Operation Manual SurgiVet V3395
3-1
Chapter 3: Setting Up the Monitor
Unpacking the Monitor and Checking the Shipment
Carefully remove the monitor and accessories from the shipping carton. Save the packing materials in
case the monitor or accessories must be shipped or stored. Compare the packing list with the
accessories received to make sure the shipment is complete.
Turning Alarm and Alert Tones On and Off
When the monitor is turned on, the alarm and alert tones are silenced for two minutes. The Alarm
Silence LED, on the LCD keypad, flashes during the two minute time-out.
•
To silence the alarm and alert tones indefinitely: Press and hold
B
for about three
seconds; the Alarm Silence LED lights steady.
NOTE: To comply with government requirements for patient monitoring, the indefinite
alarm and alert tone silence feature may not be available in monitors shipped to your
country.
•
To silence the alarm and alert tones for two minutes: Momentarily press
B
; the
Alarm Silence indicator flashes. If tones are already silenced, press
B
twice (the
first press cancels alarm silenced; the second press silences the alarms for two
minutes).
•
To cancel either two minute or indefinite alarm silence and enable alarm and alert
tones: Momentarily press
B
; the Alarm Silence LED turns off.
Working With Menus
ALARMS
VOLUME
OXIMETER
RESPIRATION
TEMPERATURE
TRENDS
PRINTER
[EXIT]
SPO2 SENSOR!
85
98
101.2 F
15
FIGURE 3.1: MENU