8.4 Correction Factor
When calibrating TCS-1000 suspended solids electrode using turbidity standard solution,
a correction factor needs to be applied to adjust the measured values for suspended solids
measurement values.
Once “Known Solution Calibration” is complete (refer to section 8.3.1.1 for TCS-1000
suspended solids electrode calibration procedure), correction factor adjustment screen will
follow. Users may also directly enter the correction factor adjustment screen from setup
mode.
Known Solution Calibration
Display Calibration Result
Rinse the sensor with distilled water
and place into known Suspended
Solids standard solution. Wait until the
value is stable, then adjust the F-factor
using
____
and until the main
measurement value reflects the target
Suspended Solids value.
Press to confirm it.
Press to confirm it.
Calibration results
overview
Press or to
move between pages.