7.13 RS-485 Communication
After entering setup mode, select “RS-485” and press to configure RS-485 communication.
According to the Modbus protocol, set the transmitting mode, parity, baud rate, stop bit, and ID
number. For Modbus protocol details, please refer to chapter 9. When under hold status, the
measurement signal output maintains the last output value before hold.
Press
or to
select RTU or ASCII mode.
Enter “Sample Measurement Average (Digital Filter)” Setup
Press
or to set the ID
number of the transmitter. The
valid value is from 1 to 247.
Press
or to
select stop bit as 1 bit or 2
bit.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press
or to
select baud rate from 2400,
4800, 9600 or 19200.
Press
or to
select parity check from
Even, Odd or None.