7.10 Clean
After entering setup mode, select “Clean” and press to activate and configure cleaning
settings. Select “Auto” to turn on the function as automatic. Set the timer for the clean function
to automatically turn cleaning on or off, and then set the Hysteresis value (Hys.). The
relationship of these settings can be seen in the graph provided below.
Note: When the clean function is turned on, if any value is set to 0, the instrument will automatically
turn off this function. When the clean function is activated under measurement mode, a “Clean
Running” message will show on the top of the display. The measurement value will remain at the last
measured value before cleaning. When entering settings menu or calibration menu while the clean
function is activated, the instrument will automatically stop the cleaning action.
Press
or to select
to activate cleaning test. If
turned
ON,
the
Clean
function will begin and the
“WASH”
indicator
will
light up.
Press
or
to select to activate
cleaning. If turned
off, the menu will
proceed to setup of
Analog
on
off
on
off
on
off
off
on
on
off
T
on
T
hys
T
off
T
on
T
hys
T
off
T
off
Clean Timer Control
Relay
Contact
Timer
Settings
Measure-
ment
Enter “Analog” Setup
Press
or
to adjust the
cleaning OFF
duration.
Press to
confirm minutes
and proceed to
seconds.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press or
to adjust Clean hysteresis
(Hys.) time.
Press
or
to adjust the
cleaning ON
duration.
Press to
confirm minutes
and proceed to
seconds.
29