9.2 MODBUS name and address table
Function Code
:
03H, 06, 10H Modbus response (setup parameter)
Logic
address
Item
Number
of Byte
Informati
on type
Description of
data transmission
Default value
Note
0001H Equipment’s
ID
2 USHORT
1-247
1
0002H
Transmitter
model
6 USHORT
ASCII
Code
PC3110
0005H
Communication
protocol
2 USHORT
0: RTU
0
1: ASCII
0006H
Serial
transmission
speed (Baud rate)
2 USHORT
0: 2400
3
1: 4800
2: 9600
3: 19200
0007H Parity
2
USHORT
0: None
1
1: Even
2: Odd
0008H
Real-time clock*
12
USHORT
Second
2012-01-01,
00:00:00
0009H USHORT
Minute
000AH USHORT
Hour
000BH USHORT
Day
000CH USHORT
Month
000DH USHORT
Year
000EH
Code setting*
2
USHORT
Code setting
1111
000FH
Temperature
mode*
2 USHORT
0: MTC
0
1: PTC
2: NTC
0010H
Clean relay*
2
USHORT
0: OFF
0
USHORT
1: AUTO
0011H 2
USHORT
ON.S:
0-5999
0
Second
0012H 2
USHORT
OFF.H:
0-999
0
Hour
0013H 2
USHORT
OFF.M:
0-59
0
Minute
0014H 2
USHORT
Hys.S:
0-9999
0
Second
0015H
Relay 1 *
2 USHORT
0: OFF
1
1: AUTO
0016H 2
USHORT
0: AUTO
0
1: Lo
0017H 4
FLOAT
SP1
10.00pH/1000mV
Data
54