7.13 RS485 communication
Enter setup of RS485 communications. According to the Modbus protocol, set the transmitting
mode, parity, baud rate, and ID number. About the detail of Modbus protocol, please refer to Ch9. If
under hold status, the measurement signal output maintains the last output value before hold status.
Press
or to
select RTU or ASCII mode.
Enter “ Sample average of measurements (Digital Filter) ” Setup
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press
or to
select Even, or Odd, or
None for parity check.
Press
or to
select 2400, or 4800, or
9600 for baud rate.
Press
or to set
the ID number of the
transmitter. The valid value
is from 1 to 247.
36