8. Calibration
Block diagram of Calibration
Cal. Info.
TECH
NIST
:
:
:
Code
Input
Password
Preset: 1100
OFF
ON
New
Code
Setting
Cal. Mode
Multi-Cal =1
or press Cal
Calculate zero, slop, sensitivity
and determination coefficient
Return to previous /
action or end of
calibration
Temp. adjust for
first buffer
( MTC or ATC )
Reading for
first buffer
Temp. adjust for
second buffer
( MTC or ATC )
Temp. adjust for
fifth buffer
( MTC or ATC )
Same
procedure for
3rd, 4th point
Temp. adjust for
first buffer
( MTC or ATC )
Any
Reading for first
buffer and adjust
Temp. adjust for
second buffer
( MTC or ATC )
Reading for
second buffer
and adjust
Temp. adjust for
fifth buffer
( MTC or ATC )
Reading for fifth
buffer and adjust
Reading for
second buffer
Reading for
fifth buffer
Multi-Cal =2
or press Cal
Multi-Cal =2
or press Cal
Multi-Cal =1
or press Cal
Same
procedure for
3rd, 4th point
Multi-Cal =1
or press Cal
Temp. adjust for
first buffer
( MTC or ATC )
Reading for
first buffer
Temp. adjust for
second buffer
( MTC or ATC )
Temp. adjust for
third buffer
( MTC or ATC )
Reading for
second buffer
Reading for
third buffer
Multi-Cal =2
or press Cal
Multi-Cal =4
or press Cal
Multi-Cal =4
or press Cal
Return
Time
Input
Auto
Manual
Exit
Define
Reading for
first buffer
Reading for
second buffer
Reading for
fifth buffer
Multi-Cal =2
or press Cal
Multi-Cal =1
or press Cal
Same
procedure for
3rd, 4th point
Set table
pH/Temp.
data
Start Cal.
Temp. adjust for
first buffer
( MTC or ATC )
Select first
buffer set
Temp. adjust for
second buffer
( MTC or ATC )
Select
second
buffer set
Temp. adjust for
fifth buffer
( MTC or ATC )
Select the
fifth
buffer set
Multi-Cal =2
or press Cal
42