8.3.8 Define calibration
Define mode has pre-set 5 buffer solution pH/Temperature table. Users may modify
the data sheet according to the buffer solutions set they use. The table can be modified
and saved in the memory of the transmitter.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press or to
select table modification
or execute preset
calibration procedure.
Press or to modify the
pH value of buffer 1 data sheet
under different temperature.
Select “Next Page”, and then
press to go to next buffer
data sheet.
Press or to modify the
pH value of buffer 2 data sheet
under different temperature.
Select “Next Page”, and then
press to go to next buffer
data sheet.
Press or to modify the
pH value of buffer 5 data sheet
under different temperature.
Select “Exit”, and then press
to finish the setting.
Press or to modify the
pH value of buffer 2 data sheet
under different temperature.
Select “Next Page”, and then
press to go to next buffer
data sheet.