7.6 Product Adjustment
Enter setup of “Product Adj.”, make the sample reading modifications. Users are
allowed to make sample reading adjustment without taking out the sensor and making
calibration. Utilize the function to adjust the field measurement as same as the lab
measurement to eliminate the doubt of measurement error. Under normal measurement
display mode, there is a PDT sign on top of the pH unit. (Please see Ch 6.4)
Enter ”Temperature” Setup
You may compare the pH reading with that of taking
the sample out and testing in the lab. Press or
to adjust the pH value. You may also adjust the ORP
reading under ORP mode through the same procedure.
Press to confirm it.
Press to confirm it.
29