8.3 Zero-Point Calibration
The inductive sensor connected with the transmitter needs to make a zero point calibration in the
air. Before calibration, it is necessary to make sure that the surface of the inductive sensor is dry
and clean and also in the air. After selecting Zero point calibration and then pressing Enter, the
transmitter will directly calibrate zero point and then display the value.
Press to confirm it.
As the main value is stable and auxiliary value is 0, which
means the calibration has been finished well, and then press
to confirm it and see the calibration information.